Crosstalk triggered safety standby pacing associated with an improperly seated ventricular lead.
Клинический случайСтатьяПеревод пока не подготовлен
Аннотация
Ventricular Safety Standby (VSS) was designed to prevent the potentially serious consequences of crosstalk in many dual-chamber pulse generators. In this report, we describe a patient in whom the VSS was activated in a special way. When the VSS presents during bipolar sensing configuration, the clinician should notice the proper electrical contact between leads and pulse generator.
Для цитирования: Sun Qi, Chen Keping, Zhang Shu. Crosstalk triggered safety standby pacing associated with an improperly seated ventricular lead.. Europace : European pacing, arrhythmias, and cardiac electrophysiology : journal of the working groups on cardiac pacing, arrhythmias, and cardiac cellular electrophysiology of the European Society of Cardiology. 2007. DOI: 10.1093/europace/eum250.