Performance on clinical outcomes, activities of daily living and user experience on head-mounted displays for people with vision impairment.

van der Aa Hilde P A, Garcia-Piña Fernanda, van Nispen Ruth M A, Hoogland Jeroen, Roberts Calvin, Seiple William

СтатьяПоддержка исследования: вне правительства СШАПеревод пока не подготовлен

Аннотация

PURPOSE: To compare the objective performance, acceptance and usability of head-mounted displays (HMDs) to provide evidence-based data that could be used to increase the efficiency of device referrals based upon a person's vision loss and functional needs.

METHODS

A cross-sectional, counterbalanced, individually controlled crossover study was performed on 15 adults with various eye conditions. Performance was measured when using four HMDs: eSight4, Eyedaptic EYE3, Eyedaptic EYE4 and IrisVision Inspire. Performance on clinical visual acuity tests and contrast were assessed, as well as vision-related activities of daily living (ADL) which were divided into three categories: Reading, Searching & Identifying and Eye-hand Coordination. User-experience was also assessed. Logistic regression analyses, Friedman one-way repeated measure analyses of variance by ranks and multivariate permutation testing were used for analysis.

RESULTS

There was a significant improvement in visual acuity when using all devices. For contrast tasks, only the eSight4 and Eyedaptic EYE3 improved performance relative to baseline. For most Reading and Searching & Identifying tasks, the odds of being able to perform the tasks were significantly higher while using the devices. However, the actual performance with most devices (e.g., number of words read or reading speed) did not improve significantly over baseline for most tasks. For the Eye-hand Coordination tasks, participants performed equivalent to or significantly poorer than baseline when using the devices. No demographic or clinical predictors of outcomes were identified. Participants expressed dissatisfaction with the devices' effectiveness, acceptability and usability.

CONCLUSIONS

While performance on clinical tests was better when using the devices, performance on most real-world ADLs was equal to or worse than baseline. No single device improved performance on all tasks, and performance on any one task was not improved with all the devices. The overall dissatisfaction with the devices paralleled the lack of objective improvement in the performance of real-world tasks.

Для цитирования: van der Aa Hilde P A, Garcia-Piña Fernanda, van Nispen Ruth M A, Hoogland Jeroen, Roberts Calvin, Seiple William. Performance on clinical outcomes, activities of daily living and user experience on head-mounted displays for people with vision impairment.. Ophthalmic & physiological optics : the journal of the British College of Ophthalmic Opticians (Optometrists). 2024. DOI: 10.1111/opo.13329.

Предметные рубрики MeSH

Подготовка и перевод публикации
Обработка статьи · этапы, ошибки и управление
Загрузка состояния…

История · было → стало

Состояние обновляется каждые 5 секунд, пока блок открыт. Процент показывается только там, где известен объём работы. Пауза останавливает продолжение этапа; текущий запрос к источнику может завершиться.

Тип публикации указан по данным источника. Уровень доказательности автоматически не присваивается.