DOI 10.1016/j.hrthm.2024.07.006

Медицинская библиотека · Heart rhythm · 2024

Performance of the quadripolar CRT-D system: Five-year results from the Quadripolar Pacing Post-Approval Study.

Corbisiero Raffaele, Cao Michael, Muller David, Lee Kwangdeok, Martin David, QUAD-PAS investigators

СтатьяМногоцентровое исследованиеНаблюдательное исследованиеПоддержка исследования: вне правительства СШАПеревод пока не подготовлен

Аннотация

BACKGROUND

Cardiac resynchronization therapy (CRT) is associated with challenges such as elevated capture thresholds, diaphragmatic stimulation, and lead instability.

OBJECTIVE

This study aimed to assess the long-term safety and efficacy of the quadripolar CRT-defibrillator (CRT-D) device system with the Quartet 1458Q left ventricular (LV) lead in a CRT-indicated population observed for 5 years and to evaluate all-cause mortality and impact of baseline characteristics on survival through 5 years.

METHODS

Patients indicated for a CRT-D system were observed every 6 months after implantation for 5 years, and device performance and adverse events were assessed at each visit. The 3 primary end points were freedom from quadripolar CRT-D system-related complications through 5 years, freedom from Quartet 1458Q LV lead-related complications through 5 years, and mean programmed pacing capture threshold at 5 years.

RESULTS

The study enrolled 1970 participants at 71 sites. The quadripolar CRT-D system was successfully implanted in 97.2% of participants. Freedom from quadripolar CRT-D device system-related complications through 5 years was 89.7%. Freedom from Quartet 1458Q LV lead-related complications through 5 years was 95.7%; 3.49% of participants had LV lead-related complications, and an overall LV lead complication rate was 0.0122 event per patient-year. The mean LV pacing capture threshold was 1.52 ± 1.01 V at 5 years. The 5-year survival rate was 67.4%.

CONCLUSION

The quadripolar CRT-D system with the Quartet 1458Q LV lead exhibited low rates of complications and stable electrical performance through 5 years of follow-up and suggested a higher 5-year survival rate compared with traditional CRT systems.

Для цитирования: Corbisiero Raffaele, Cao Michael, Muller David, Lee Kwangdeok, Martin David, QUAD-PAS investigators. Performance of the quadripolar CRT-D system: Five-year results from the Quadripolar Pacing Post-Approval Study.. Heart rhythm. 2024. DOI: 10.1016/j.hrthm.2024.07.006.