Comparison of High-tensile Suture and 16-gauge Wire for Prophylactic Fixation in Total Hip Arthroplasty: A Biomechanical Study.

Guarin Perez Sergio F, Zhao Gongyin, Tsukamoto Ichiro, Labott Joshua R, Restrepo Diego J, Tai Ta-Wei, Hooke Alexander W, Zhao Chunfeng, Sierra Rafael J

СтатьяПеревод пока не подготовлен

Аннотация

BACKGROUND

Prophylactic cerclage fixation may reduce fracture risk after total hip arthroplasty. Traditionally performed with wire or cables, high-tensile suture is a novel alternative not previously evaluated. This study compared peak torque to failure of cemented and cementless femoral stems using either 16-gauge wire or high-tensile suture in a synthetic femur model.

METHODS

Sixteen collarless triple-taper (TT) stems (8 calcar wire, 8 calcar suture) and 27 taper-slip (TS) stems were tested: 7 each with calcar suture, diaphyseal wire, and diaphyseal suture, and 6 with calcar wire (one excluded). Stems were press-fit or cemented into composite femora. Fixation was applied above the lesser trochanter (calcar) or 2 cm below (diaphyseal). Sutures were tensioned to 80 N. A servohydraulic system applied a 2000 N load with 1°/s internal rotation until failure. Power analysis supported 7 specimens per group to detect a 30 Nm difference. Two-sample and noninferiority t-tests were performed.

RESULTS

For TT stems, peak torque was 92.7 ± 11.4 Nm with calcar wire and 102.6 ± 41.4 Nm with calcar suture; suture was noninferior (P = .038). For TS stems, torque was 148.2 ± 38.6 Nm (calcar wire), 118.3 ± 30.2 Nm (calcar suture), 164.9 ± 21.5 Nm (diaphyseal wire), and 159.4 ± 30.0 Nm (diaphyseal suture). Noninferiority in TS stems was not established. Diaphyseal suture exceeded calcar suture (P = .038).

CONCLUSIONS

High-tensile suture was comparable to wire for calcar fixation in collarless TT stems. In TS stems, noninferiority was not demonstrated. Suture constructs may offer practical advantages and warrant further study.

Для цитирования: Guarin Perez Sergio F, Zhao Gongyin, Tsukamoto Ichiro, Labott Joshua R, Restrepo Diego J, Tai Ta-Wei, Hooke Alexander W, Zhao Chunfeng, Sierra Rafael J. Comparison of High-tensile Suture and 16-gauge Wire for Prophylactic Fixation in Total Hip Arthroplasty: A Biomechanical Study.. Arthroplasty today. 2026. DOI: 10.1016/j.artd.2026.101966.
Подготовка и перевод публикации
Обработка статьи · этапы, ошибки и управление
Загрузка состояния…

История · было → стало

Состояние обновляется каждые 5 секунд, пока блок открыт. Процент показывается только там, где известен объём работы. Пауза останавливает продолжение этапа; текущий запрос к источнику может завершиться.

Тип публикации указан по данным источника. Уровень доказательности автоматически не присваивается.