Quantifying the Transmit-Power Threshold for Reliable Passive Detuning of Inductively Coupled Wireless Resonators in MRI: A Phantom Study.

Lu Ming, Jiang Xiaoyu, Ma Yajun, Qian Yurui, Liao Congyu, Yang Yang, Gore John C, Yan Xinqiang

СтатьяПеревод пока не подготовлен

Аннотация

This work develops and validates a practical method for quantifying the transmit-power threshold at which passive detuning of inductively coupled wireless resonators becomes reliable in MRI. All experiments were performed using phantoms. A near/far spatial signal ratio is computed from a series of gradient-echo (GRE) images acquired at different nominal flip angles. This ratio reveals ineffective detuning as a characteristic L-shaped curve whose inflection point marks the detuning threshold. The method was validated experimentally with wireless resonators incorporating different numbers of crossed diode pairs, at multiple static field strengths (3 and 7T), and with different scanner vendors. At 3T, the method produced consistent B 1 + thresholds across Philips and GE scanners despite their different flip-angle thresholds. The proposed method provides a practical, straightforward tool for acceptance testing and quality assurance of wireless resonator safety in both clinical and research MRI settings.

Для цитирования: Lu Ming, Jiang Xiaoyu, Ma Yajun, Qian Yurui, Liao Congyu, Yang Yang, Gore John C, Yan Xinqiang. Quantifying the Transmit-Power Threshold for Reliable Passive Detuning of Inductively Coupled Wireless Resonators in MRI: A Phantom Study.. NMR in biomedicine. 2026. DOI: 10.1002/nbm.70408.

Предметные рубрики MeSH

Подготовка и перевод публикации
Обработка статьи · этапы, ошибки и управление
Загрузка состояния…

История · было → стало

Состояние обновляется каждые 5 секунд, пока блок открыт. Процент показывается только там, где известен объём работы. Пауза останавливает продолжение этапа; текущий запрос к источнику может завершиться.

Тип публикации указан по данным источника. Уровень доказательности автоматически не присваивается.