Performance Evaluation of Automated Erythrocyte Sedimentation Rate (ESR) Analyzers in a Multicentric Study.

Tomassetti Flaminia, Calabrese Cinzia, Bertani Fabio, Cennamo Michele, Diamanti Daniela, Giovannelli Alfredo, Guerranti Roberto, Leoncini Roberto, Lorubbio Maria, Ognibene Agostino, Nicolai Eleonora, Pelagalli Martina, Pieroni Carolina, Bernardini Sergio, Pieri Massimo

СтатьяПеревод пока не подготовлен

Аннотация

BACKGROUND

Erythrocyte Sedimentation Rate (ESR) is an easy test used to diagnose and monitor inflammatory and infectious diseases. The aim of this study was the evaluation of the performance of three ESR automated analyzers, VES-MATIC 5, CUBE 30 TOUCH, and MINI-CUBE, involving four Italian polyclinics in Rome, Siena, Como, and Arezzo, as well as inter-site variability assessment to detect possible device-dependent and operator-dependent influences.

METHODS

Accuracy analysis was carried out by analyzing the same samples with all three instruments and comparing them with the Westergren method. Precision was assessed with quality control material through intra-run and inter-run precision. Repeatability was estimated by reanalyzing fresh blood samples belonging to three ESR ranges (low, intermediate, and high) six times.

RESULTS

The results showed a strong correlation (Spearman coefficients R2) between the manual method and VES-MATIC 5 (0.978), CUBE 30 TOUCH (0.981), and MINI-CUBE (0.974). The accuracy of all clinics was excellent, with coefficients of variation (CVs) of less than 10% for all instruments. Repeatability confirmed an excellent level for all ESR ranges, with CVs below 10%.

CONCLUSIONS

The study proved that all three automated instruments offer optimal performance for accuracy and precision and are suitable for both large and small facilities without influences of the laboratory environment.

Для цитирования: Tomassetti Flaminia, Calabrese Cinzia, Bertani Fabio, Cennamo Michele, Diamanti Daniela, Giovannelli Alfredo, Guerranti Roberto, Leoncini Roberto, Lorubbio Maria, Ognibene Agostino, Nicolai Eleonora, Pelagalli Martina, Pieroni Carolina, Bernardini Sergio, Pieri Massimo. Performance Evaluation of Automated Erythrocyte Sedimentation Rate (ESR) Analyzers in a Multicentric Study.. Diagnostics (Basel, Switzerland). 2024. DOI: 10.3390/diagnostics14182011.
Подготовка и перевод публикации
Обработка статьи · этапы, ошибки и управление
Загрузка состояния…

История · было → стало

Состояние обновляется каждые 5 секунд, пока блок открыт. Процент показывается только там, где известен объём работы. Пауза останавливает продолжение этапа; текущий запрос к источнику может завершиться.

Тип публикации указан по данным источника. Уровень доказательности автоматически не присваивается.