Poorer outcomes associated with more invasive lead management strategies for Abbott Riata® leads: a large, multicenter experience.

Mar Philip L, Tsai Christina, Golemi Lolita, Bedi Rohil, Khan Ali, Kabra Rajesh, Atkins Donita, Bhakta Deepak, Lakkireddy Dhanunjaya, Gopinathannair Rakesh

СтатьяМногоцентровое исследованиеСравнительное исследованиеПеревод пока не подготовлен

Аннотация

BACKGROUND

Over 100,000 Abbott Riata® were implanted in the United States before they were recalled in 2010. There are still a significant number of Abbott Riata® leads in use, and it is unclear how these leads should be managed at the time of generator change or lead malfunction. Although data comparing both Sprint Fidelis® and Abbott Riata® leads in this setting is available, there are no multicenter comparative studies of outcomes for various lead management strategies, including lead extraction (LE), lead abandonment/revision (LA), and generator change (GC) only at the time of device at elective replacement interval (ERI) for Abbott Riata® leads.

METHODS

A retrospective, multicenter study was undertaken to compare short-term outcomes (major complications-MC, death, extended or re-hospitalizations within 60 days-RH, lead malfunction-LM) and total outcomes (short-term outcomes & lead malfunction during follow-up) of patients with Riata® leads undergoing LE, LA, or GC.

RESULTS

152 patients (65 ± 13 years, 68% male) were followed for a mean 33 ± 30 months following intervention. Out of 166 procedures, 13 patients underwent LE, 16 patients underwent LA, and 137 patients underwent GC. There was 1 major complication in each group, yielding an event rate of 7.7% for LE, 6.3% for LA, and 0.7% for GC cohorts. There were significantly more short-term and total adverse outcomes in the group of patients getting LE and LA versus GC only (38.5% & 31.3% vs 7.3%, P < 0.001). Total Riata® lead dwell time follow-up was 17,067 months. A total of 3 Riata® lead malfunctions were noted during long-term follow-up. Inappropriate shocks were similar between LE 7.7% (1/13), LA 6.3% (1/16). and GC 11.0% (4/136); P = 0.57.

CONCLUSIONS

There were more short-term and total adverse outcomes in more invasive management strategies (LE and LA) versus GC alone. The failure rate of Riata® leads was substantially lower compared to previous reports. Therefore, we recommend only performing battery exchange when a device with an active Riata® lead is at ERI, unless there is malfunction of the Riata® lead noted on testing.

Для цитирования: Mar Philip L, Tsai Christina, Golemi Lolita, Bedi Rohil, Khan Ali, Kabra Rajesh, Atkins Donita, Bhakta Deepak, Lakkireddy Dhanunjaya, Gopinathannair Rakesh. Poorer outcomes associated with more invasive lead management strategies for Abbott Riata® leads: a large, multicenter experience.. Journal of interventional cardiac electrophysiology : an international journal of arrhythmias and pacing. 2023. DOI: 10.1007/s10840-023-01673-y.