MedElevator Войти в исследования
← Поиск публикаций

DOI 10.1016/j.jacc.2011.04.010

Медицинская библиотека · Journal of the American College of Cardiology · 2011

Incidence of asymptomatic intracranial embolic events after pulmonary vein isolation: comparison of different atrial fibrillation ablation technologies in a multicenter study.

Herrera Siklódy Claudia, Deneke Thomas, Hocini Mélèze, Lehrmann Heiko, Shin Dong-In, Miyazaki Shinsuke, Henschke Susanne, Fluegel Peter, Schiebeling-Römer Jochen, Bansmann Paul M, Bourdias Thomas, Dousset Vincent, Haïssaguerre Michel, Arentz Thomas

Сравнительное исследованиеСтатьяМногоцентровое исследованиеПоддержка исследования: вне правительства СШАПеревод пока не подготовлен

Аннотация

OBJECTIVES

We compared the safety of different devices by screening for subclinical intracranial embolic events after pulmonary vein isolation with either conventional irrigated radiofrequency (RF) or cryoballoon or multielectrode phased RF pulmonary vein ablation catheter (PVAC).

BACKGROUND

New devices specifically designed to facilitate pulmonary vein isolation procedures have recently been introduced.

METHODS

This prospective, observational, multicenter study included patients with symptomatic atrial fibrillation referred for pulmonary vein isolation. Ablation was performed using 1 of the 3 catheters. Strict periprocedural anticoagulation, with intravenous heparin during ablation to achieve an activated clotting time >300 s, was ensured in all patients. Cerebral magnetic resonance imaging was performed before and after ablation.

RESULTS

Seventy-four patients were included in the study: 27 in the irrigated RF group, 23 in the cryoballoon group, and 24 in the PVAC group. Total procedure times were 198 ± 50 min, 174 ± 35 min, and 124 ± 32 min, respectively (p < 0.001 for PVAC vs. irrigated RF and cryoballoon). Findings on neurological examination were normal in all patients before and after ablation. Post-procedure magnetic resonance imaging detected a single new embolic lesion in 2 of 27 patients in the irrigated RF group (7.4%) and in 1 of 23 in the cryoballoon group (4.3%). However, in the PVAC group 9 of 24 patients (37.5%) demonstrated 2.7 ± 1.3 new lesions each (p = 0.003 for the presence of new embolic events among the 3 groups).

CONCLUSIONS

The PVAC is associated with a significantly higher incidence of subclinical intracranial embolic events. Further study of the causes and significance of these emboli is required to determine the safety of the PVAC.

Для цитирования: Herrera Siklódy Claudia, Deneke Thomas, Hocini Mélèze, Lehrmann Heiko, Shin Dong-In, Miyazaki Shinsuke, Henschke Susanne, Fluegel Peter, Schiebeling-Römer Jochen, Bansmann Paul M, Bourdias Thomas, Dousset Vincent, Haïssaguerre Michel, Arentz Thomas. Incidence of asymptomatic intracranial embolic events after pulmonary vein isolation: comparison of different atrial fibrillation ablation technologies in a multicenter study.. Journal of the American College of Cardiology. 2011. DOI: 10.1016/j.jacc.2011.04.010.

Предметные рубрики MeSH

Подготовка и перевод публикации
Обработка статьи · этапы, ошибки и управление
Загрузка состояния…

История · было → стало

Состояние обновляется каждые 5 секунд, пока блок открыт. Процент показывается только там, где известен объём работы. Пауза останавливает продолжение этапа; текущий запрос к источнику может завершиться.

Тип публикации указан по данным источника. Уровень доказательности автоматически не присваивается.