DOI 10.1272/jnms.77.306

Медицинская библиотека · Journal of Nippon Medical School = Nippon Ika Daigaku zasshi · 2010

A comparative study of sterility levels in continuous ambulatory peritoneal dialysis system.

Saito Sabine Kyoko, Kashiwagi Tetsuya, Miyahara Hideyasu, Iino Yasuhiko, Katayama Yasuo

Сравнительное исследованиеСтатьяПеревод пока не подготовлен

Аннотация

INTRODUCTION

Peritonitis remains a serious risk associated with continuous ambulatory peritoneal dialysis (CAPD), although better patient education programs and such technological advances as improved automated connecting devices have greatly decreased its incidence over the past 20 years. The automated devices have a good resistance to contamination, but they rely on an external electrical power source and are not easily portable. There has, therefore, been a need for a highly sterile nonelectric manual connecting device to complement the automated devices already in use. Such a manual device has recently been developed. We compared the level of sterility after touch contamination in this new device with levels in 2 other connecting devices: a conventional device with a manual cap (JMS Co. Ltd., Hiroshima, Japan), and a powered total containment device (JMS Co. Ltd.).

METHOD: Five bacteria frequently causing CAPD-related peritonitis (Staphylococcus aureus, Staphylococcus epidermidis, Escherichia coli, Pseudomonas aeruginosa, and Candida albicans) were separately applied to the tip of each connecting device, and peritoneal washing fluid was injected as in a routine exchange. We used a membrane filter method to determine whether the route had been contaminated by the washing fluid.

RESULTS

In the conventional device with a manual cap, 3 to 4 colony-forming units (CFUs) of S. aureus were detected in 2 of 10 drainage samples, 8 CFUs of E. coli in 1 of 10 drainage samples, and 1 CFU of E. coli in the injection fluid. In contrast, no contamination was detected in the automated connecting device or the new manual cap device.

CONCLUSION

This study confirmed that the new device has a risk of touch contamination lower than that of the conventional manual cap device and equal to that of the automated device. Being easily portable and not reliant on an external power source, the new device should be useful in various situations.

Для цитирования: Saito Sabine Kyoko, Kashiwagi Tetsuya, Miyahara Hideyasu, Iino Yasuhiko, Katayama Yasuo. A comparative study of sterility levels in continuous ambulatory peritoneal dialysis system.. Journal of Nippon Medical School = Nippon Ika Daigaku zasshi. 2010. DOI: 10.1272/jnms.77.306.

Предметные рубрики MeSH

Подготовка и перевод публикации
Обработка статьи · этапы, ошибки и управление
Загрузка состояния…

История · было → стало

Состояние обновляется каждые 5 секунд, пока блок открыт. Процент показывается только там, где известен объём работы. Пауза останавливает продолжение этапа; текущий запрос к источнику может завершиться.

Тип публикации указан по данным источника. Уровень доказательности автоматически не присваивается.