A falsely low lactate: discrepancies in lactate measurement between blood gas and core laboratory instruments due to β-hydroxybutyrate interference-a case report.
СтатьяКлинический случайНазвание и аннотация: Перевод пока не подготовлен
Обработка статьи · этапы, ошибки и управление
Загрузка состояния…
Ошибки и ограничения
История · было → стало
Состояние обновляется каждые 5 секунд, пока блок открыт. Процент показывается только там, где известен объём работы. Пауза останавливает продолжение этапа; текущий запрос к источнику может завершиться.
Alqurini Nadia M, Adhikari Bhawana, Hagrass Hoda A
Аннотация
INTRODUCTION: Laboratory tests play a crucial role in diagnosing and managing illness. Accurate interpretation, which includes evaluating false-negative and false-positive results, is essential for guiding appropriate clinical interventions and ensuring patient safety.
METHODS: We present a case involving clinically significant discrepancies in lactate measurements in a patient with acute myeloid leukemia and diabetic ketoacidosis.
RESULTS: These discrepancies were caused by interference from elevated levels of ketone bodies-specifically, β-hydroxybutyrate-which may have delayed recognition of the severity of the patient's condition.
DISCUSSION: Dilution and spike experiments suggested potential interferences in core laboratory instruments using the Trinder reaction, leading to inaccurate lactate readings.
METHODS: We present a case involving clinically significant discrepancies in lactate measurements in a patient with acute myeloid leukemia and diabetic ketoacidosis.
RESULTS: These discrepancies were caused by interference from elevated levels of ketone bodies-specifically, β-hydroxybutyrate-which may have delayed recognition of the severity of the patient's condition.
DISCUSSION: Dilution and spike experiments suggested potential interferences in core laboratory instruments using the Trinder reaction, leading to inaccurate lactate readings.