Corneal and Epithelial Thickness Mapping: Comparison of Swept-Source- and Spectral-Domain-Optical Coherence Tomography.

Georgeon Cristina, Marciano Ilanite, Cuyaubère Roxane, Sandali Otman, Bouheraoua Nacim, Borderie Vincent

СтатьяПеревод пока не подготовлен

Аннотация

OBJECTIVE

To compare the results and repeatability of the corneal thickness (CT) and epithelial thickness (ET) maps provided by Swept-Source-Optical Coherence Tomography with those of Spectral-Domain-OCT in normal eyes.

METHODS

30 normal eyes of 30 patients were assessed by 3 trained operators with SS-OCT and SD-OCT.

RESULTS

The central and minimum ET obtained with both devices were correlated: central ET, r = 0.86, p < 0.05; minimum ET, r = 0.72, p < 0.05. Compared with SD-OCT, SS-OCT tended to underestimate these figures by 1.4 and 1.9 μm on average. The central and minimum CT obtained with both devices were strongly correlated: central CT, r = 0.994, p < 0.05; minimum CT, r = 0.995, p < 0.05. SS-OCT tended to overestimate these figures by 11 and 14 μm on average. Repeatability was good for both devices with a mean coefficient of variation of measurements <6% for ET and <2% for CT. Interoperator variability (standard deviation and COV) was significantly higher for SS-OCT than for SD-OCT for all local epithelial thicknesses and significantly lower for the central CT and several local corneal thicknesses, whereas no significant differences between both technologies were found for the central and minimum ET and the minimum CT.

CONCLUSION

SS-OCT and SD-OCT provide reproducible measurements of CT and ET in normal corneas with a strong correlation between both technologies. However, both technologies are not interchangeable when the main thickness parameters (i.e., central and minimum CT and minimum ET) are used for diagnosing early keratoconus or calculating the expected residual stromal bed thickness before corneal refractive surgery or anterior lamellar keratoplasty.

Для цитирования: Georgeon Cristina, Marciano Ilanite, Cuyaubère Roxane, Sandali Otman, Bouheraoua Nacim, Borderie Vincent. Corneal and Epithelial Thickness Mapping: Comparison of Swept-Source- and Spectral-Domain-Optical Coherence Tomography.. Journal of ophthalmology. 2021. DOI: 10.1155/2021/3444083.
Подготовка и перевод публикации
Обработка статьи · этапы, ошибки и управление
Загрузка состояния…

История · было → стало

Состояние обновляется каждые 5 секунд, пока блок открыт. Процент показывается только там, где известен объём работы. Пауза останавливает продолжение этапа; текущий запрос к источнику может завершиться.

Тип публикации указан по данным источника. Уровень доказательности автоматически не присваивается.