MedElevator Войти в исследования
← Поиск публикаций

DOI 10.1259/bjr.20170615

Медицинская библиотека · The British journal of radiology · 2018

A tertiary centre experience of thoracic CT and cardiac MRI scanning in the presence of a reveal LINQ insertable cardiac monitoring system: a case series review of artefact, patient safety and data preservation.

Wiles Benedict M, Illingworth Caroline A, Couzins Michael Lg, Roberts Paul R, Harden Stephen P

СтатьяПеревод пока не подготовлен

Аннотация

OBJECTIVE

At our tertiary cardiothoracic centre, cardiac MRI and thoracic CT scans are performed in patients with implanted LINQ devices. The degree of foreign body artefact associated with the LINQ device, and its clinical importance, has not previously been assessed. A case series review was therefore performed with a simultaneous review of patient safety and data loss events, secondary to the MRI environment.

METHODS

A local database search identified LINQ device patients who underwent thoracic CT or cardiac MRI scans between March 2014 and December 2016. Images were reviewed by two radiologists, recording the presence of subcutaneous and intrathoracic artefact, and its clinical significance. Furthermore a specialist in cardiac rhythm management reviewed all LINQ data downloads undertaken before and after MRI scanning, and a search of the trust incident reporting system was performed.

RESULTS

Minor subcutaneous artefact was present on all scans. Intrathoracic artefact was observed in 25.6% of thoracic CT scans and 33.3% of cardiac MRIs; however no clinically significant artefact was observed. Device downloads were only performed by 53.8% of patients prior to their MRI scan and 56.5% after their MRI scan. No adverse patient safety or data loss events were noted.

CONCLUSION

The LINQ device does not produce clinically significant artefact, even when artefact extends into the intrathoracic space, which occurs in a third of MRIs and a quarter of CTs. MRI scanning of the LINQ device is safe with no evidence of inappropriate data loss. Advances in knowledge: This is the first published case series of CT and MRI scanning in LINQ patients and the first performed quantification of artefact related to the LINQ device.

Для цитирования: Wiles Benedict M, Illingworth Caroline A, Couzins Michael Lg, Roberts Paul R, Harden Stephen P. A tertiary centre experience of thoracic CT and cardiac MRI scanning in the presence of a reveal LINQ insertable cardiac monitoring system: a case series review of artefact, patient safety and data preservation.. The British journal of radiology. 2018. DOI: 10.1259/bjr.20170615.

Предметные рубрики MeSH

Подготовка и перевод публикации
Обработка статьи · этапы, ошибки и управление
Загрузка состояния…

История · было → стало

Состояние обновляется каждые 5 секунд, пока блок открыт. Процент показывается только там, где известен объём работы. Пауза останавливает продолжение этапа; текущий запрос к источнику может завершиться.

Тип публикации указан по данным источника. Уровень доказательности автоматически не присваивается.