MedElevator Войти в исследования
← Поиск публикаций

DOI 10.1161/circinterventions.116.004244

Медицинская библиотека · Circulation. Cardiovascular interventions · 2017

Independent Modular Filter for Embolic Protection in Carotid Stenting.

Scheinert Dierk, Reimers Bernhard, Cremonesi Alberto, Schmidt Andrej, Sievert Horst, Rohde Stefan, Schofer Joachim, Mudra Harald G, Bosiers Marc, Zeller Thomas, Pacchioni Andrea, Rosenschein Uri, WISE (Wirion Study Europe) Pivotal Trial Investigators

Клиническое исследованиеСтатьяМногоцентровое исследованиеПеревод пока не подготовлен

Аннотация

BACKGROUND

Embolic protection during carotid artery stenting reduces the rate of thromboembolic events. The Wirion Embolic Protection System is used to deploy an independent distal filter using any 0.014″ guidewire. WISE study (Wirion Study Europe) evaluated the safety and performance of Wirion Embolic Protection System in patients undergoing carotid artery stenting.

METHODS AND

RESULTS

A prospective, multicenter, nonrandomized, open-label, single-arm study of carotid artery stenting in high surgical risk patients was performed. The primary end point, a composite of death, stroke, and myocardial infarction at 30 days, was compared with performance goal derived from historical controls. Secondary end points were components of the primary end point and the device, angiographic, procedural, and clinical success rates. Preplanned interim analysis was performed on the first 120 patients. At interim analysis, the primary end point was significantly lower for the Wirion Embolic Protection System group, compared with historical data (3.3% versus 6.3%, respectively; P value =0.0008). Analysis of primary end point components in the WISE group, compared with the historical control group, shows numerically lower mortality (0% versus 1.7%, respectively; P=0.21), stroke (2.5% versus 4.6%, respectively; P=0.18), and myocardial infarction (0.8% versus 1.5%, respectively; P=0.50). Device, angiographic, procedural, and clinical success was achieved in 99.2%, 99.1%, 98.3%, and 96.6% of cases, respectively.

CONCLUSIONS

The data suggest that independent modular filter use in carotid artery stenting in high surgical risk patients is safe and effective. The outcomes suggest that use of an independent modular filter may be associated with a lower rate of embolic complications associated with carotid stent placement.

CLINICAL TRIAL REGISTRATION: URL: http://www.clinicaltrials.gov. Unique identifier: NCT01783639.

Для цитирования: Scheinert Dierk, Reimers Bernhard, Cremonesi Alberto, Schmidt Andrej, Sievert Horst, Rohde Stefan, Schofer Joachim, Mudra Harald G, Bosiers Marc, Zeller Thomas, Pacchioni Andrea, Rosenschein Uri, WISE (Wirion Study Europe) Pivotal Trial Investigators. Independent Modular Filter for Embolic Protection in Carotid Stenting.. Circulation. Cardiovascular interventions. 2017. DOI: 10.1161/circinterventions.116.004244.

Предметные рубрики MeSH

Подготовка и перевод публикации
Обработка статьи · этапы, ошибки и управление
Загрузка состояния…

История · было → стало

Состояние обновляется каждые 5 секунд, пока блок открыт. Процент показывается только там, где известен объём работы. Пауза останавливает продолжение этапа; текущий запрос к источнику может завершиться.

Тип публикации указан по данным источника. Уровень доказательности автоматически не присваивается.