MedElevator Войти в исследования
← Поиск публикаций

DOI 10.1097/ta.0b013e31829a2c43

Медицинская библиотека · The journal of trauma and acute care surgery · 2013

Mechanical complications of intertrochanteric hip fractures treated with trochanteric femoral nails.

Liu Wanjun, Zhou Dongsheng, Liu Fang, Weaver Michael J, Vrahas Mark S

СтатьяПеревод пока не подготовлен

Аннотация

BACKGROUND

The Trochanteric Fixation Nail System (TFN) was designed with a helical blade to improve resistance to cutout. We evaluated the incidence, modes, and predictors of failure for intertrochanteric hip fractures treated with this device.

METHODS

A retrospective review of 341 intertrochanteric hip fractures treated with the TFN was conducted. Fractures were classified from the injury films using the Arbeitsgemeinschaft für Osteosynthesefragen/Orthopedic Trauma Association (AO/OTA) classification, and quality of reduction and hardware placement was evaluated on the initial postoperative films.

RESULTS

Two hundred twenty-three patients met inclusion criteria. The overall rate of mechanical complications was 20.5%. Twenty-one patients (9.4%) had excessive lateral migration of the helical blade (≥10 mm). Fifteen patients (6.7%) had blade migration in the head, including 7 patients (3.1%) with typical cutout and 8 patients (3.6%) with medial perforation without loss of reduction. Three patients (1.3%) sustained a femoral shaft fracture at the tip of the nail. The quality of calcar reduction was significantly predictive of all modes of failure (p < 0.05), except femoral shaft fracture at the nail tip.

CONCLUSION

The most common mechanical complication was lateral migration of the helical blade; however, this rarely led to a clinical problem. Medial migration of the blade in the head without loss of reduction was the second most frequent complication. We have identified a new complication associated with the TFN, that is, medial perforation without loss of reduction. Blade migration in the head was less frequent when blades were placed inferiorly with a tip-apex distance of at least 15 mm.

Для цитирования: Liu Wanjun, Zhou Dongsheng, Liu Fang, Weaver Michael J, Vrahas Mark S. Mechanical complications of intertrochanteric hip fractures treated with trochanteric femoral nails.. The journal of trauma and acute care surgery. 2013. DOI: 10.1097/ta.0b013e31829a2c43.

Предметные рубрики MeSH

Подготовка и перевод публикации
Обработка статьи · этапы, ошибки и управление
Загрузка состояния…

История · было → стало

Состояние обновляется каждые 5 секунд, пока блок открыт. Процент показывается только там, где известен объём работы. Пауза останавливает продолжение этапа; текущий запрос к источнику может завершиться.

Тип публикации указан по данным источника. Уровень доказательности автоматически не присваивается.