DOI 10.1097/ftd.0000000000000133

Медицинская библиотека · Therapeutic drug monitoring · 2015

Evaluation of QMS everolimus assay using Indiko analyzer: comparison with an ultra-performance liquid chromatography-tandem mass spectrometry method.

Buthiau Delphine, Bargnoux Anne-Sophie, Badiou Stéphanie, Sutra Thibault, Dupuy Anne-Marie, Pageaux Georges-Philippe, Mourad Georges, Cristol Jean-Paul

Сравнительное исследованиеСтатьяПоддержка исследования: вне правительства СШАПеревод пока не подготовлен

Аннотация

BACKGROUND

A new particle-enhanced turbidimetric immunoassay, Quantitative Microsphere System (QMS) everolimus, was evaluated using an Indiko analyzer (Thermo Fisher Scientific).

METHODS

Analytical performances (imprecision, linearity, limit of detection, and limit of quantification) of this new immunoassay were evaluated. The method was compared with an ultra-performance liquid chromatography-tandem mass spectrometry (UPLC/MS/MS) method and with the previously available Innofluor Certican immunoassay on 120 whole-blood samples from 74 transplant recipients. Passing-Bablok regression and Bland-Altman plot were used for method comparisons.

RESULTS

Within- and between-day coefficients of variation were <10% at mean levels of 4.3, 8.2, and 15.9 ng/mL, respectively. This assay could be used at everolimus concentrations between 1.7 and 20 ng/mL. We obtained a Passing-Bablok regression of y = 0.99 × -0.15 (r = 0.95) when comparing with the UPLC/MS/MS method and of y = 0.81 × -0.09 (r = 0.94) when comparing with Innofluor assay. The mean and limits of agreement (mean ± 1.96 SD) of the difference between QMS everolimus and UPLC/MS/MS were -0.13 (from -2.0 to 1.74) ng/mL and were -1.19 (from -3.39 to 1.01) ng/mL between QMS everolimus and Innofluor.

CONCLUSIONS

The QMS everolimus method on Indiko analyzer demonstrated reliable and reproducible performances allowing routine monitoring in transplant patients.

Для цитирования: Buthiau Delphine, Bargnoux Anne-Sophie, Badiou Stéphanie, Sutra Thibault, Dupuy Anne-Marie, Pageaux Georges-Philippe, Mourad Georges, Cristol Jean-Paul. Evaluation of QMS everolimus assay using Indiko analyzer: comparison with an ultra-performance liquid chromatography-tandem mass spectrometry method.. Therapeutic drug monitoring. 2015. DOI: 10.1097/ftd.0000000000000133.

Предметные рубрики MeSH

Подготовка и перевод публикации
Обработка статьи · этапы, ошибки и управление
Загрузка состояния…

История · было → стало

Состояние обновляется каждые 5 секунд, пока блок открыт. Процент показывается только там, где известен объём работы. Пауза останавливает продолжение этапа; текущий запрос к источнику может завершиться.

Тип публикации указан по данным источника. Уровень доказательности автоматически не присваивается.