MedElevator Войти в исследования
← Поиск публикаций

DOI 10.1186/s12890-023-02358-y

Медицинская библиотека · BMC pulmonary medicine · 2023

Evaluation of a transbronchial cryoprobe for the ablation of pulmonary nodules: an in vitro pilot study.

Hammer Dániel, Büdi Lilla, Nagy Ádám, Varga Rita, Horváth Péter

СтатьяПеревод пока не подготовлен

Аннотация

BACKGROUND

Ablation of malignant pulmonary nodules is a novel therapeutic option for patients who cannot undergo surgery. Current transthoracic approaches cause pneumothorax and/or bleeding in a significant number of cases.

OBJECTIVE

Our purpose with this study was to evaluate cryoablation under in vitro conditions with a commercially available cryosurgery system.

METHODS

We used ballistic gelatin to model the thermal conduction of lung tissue. The cryoprobe was inserted in the ballistic gelatin with two thermal sensors, they were placed 0.5 cm and 1.0. cm from the probe, respectively, temperature was measured on both sides. We used single-, double- and triple-freeze protocols to see if we could freeze it to -20 °C.

RESULTS

We achieved - 18.6 ± 3.26 °C on the closer sensor (sensor 1) and - 3.7 ± 4.61 °C on the sensor further away (sensor 2) after 15 min using the single-freeze protocol. Using the dual-freeze protocol, we achieved - 23.2 ± 2,23 °C on sensor 1 and - 16.5 ± 2.82 °C on sensor 2. With the triple-freeze protocol we obtained - 23.5 ± 2.38 °C on sensor 1 and - 19.05 ± 3.22 °C on sensor 2.

CONCLUSION

With dual-freeze, values above - 20 °C were achieved using nearer sensor data, but a plateau phase occurred as with continuous freezing. Using triple freeze, we reached - 20 °C at a distance of 0.5 cm from the probe, but not at 1 cm; therefore, we did not expand the diameter of the predicted necrosis zone.

Для цитирования: Hammer Dániel, Büdi Lilla, Nagy Ádám, Varga Rita, Horváth Péter. Evaluation of a transbronchial cryoprobe for the ablation of pulmonary nodules: an in vitro pilot study.. BMC pulmonary medicine. 2023. DOI: 10.1186/s12890-023-02358-y.

Предметные рубрики MeSH

Подготовка и перевод публикации
Обработка статьи · этапы, ошибки и управление
Загрузка состояния…

История · было → стало

Состояние обновляется каждые 5 секунд, пока блок открыт. Процент показывается только там, где известен объём работы. Пауза останавливает продолжение этапа; текущий запрос к источнику может завершиться.

Тип публикации указан по данным источника. Уровень доказательности автоматически не присваивается.