Neither the accuracy nor the precision of thermal dilution cardiac output measurements is altered by acute tricuspid regurgitation in pigs.

Buffington Charles W, Nystrom Elisabet U M

СтатьяПеревод пока не подготовлен

Аннотация

UNLABELLED: Whether measurement of cardiac output using the thermal dilution technique (TDCO) is valid in the presence of tricuspid regurgitation (TR) is controversial. We assessed the accuracy and precision of the technique in pigs by comparison with data from an electromagnetic flowmeter on the aorta (EMCO). TR was created with sutures that immobilized the free-wall leaflets of the tricuspid valve, and cardiac output was adjusted with dobutamine to give values comparable to control measurements. TR reduced forward stroke volume from 17.2 to 12.6 mL/beat and caused the right atrium to dilate and pulse in synchrony with the right ventricle. Acute TR did not affect the linear regression relation between TDCO and EMCO and did not alter the correlation coefficient (r = 0.94 during both control and TR). These data demonstrate that acute TR does not affect the accuracy or precision of TDCO in pigs.

IMPLICATIONS: Cardiac output is a valuable measurement that guides the medical care of patients with heart and lung disease. This study demonstrates that the thermal dilution technique of determining cardiac output is valid when acute tricuspid valve regurgitation is present in pigs.

Для цитирования: Buffington Charles W, Nystrom Elisabet U M. Neither the accuracy nor the precision of thermal dilution cardiac output measurements is altered by acute tricuspid regurgitation in pigs.. Anesthesia and analgesia. 2004. DOI: 10.1213/01.ane.0000105923.09732.93.

Предметные рубрики MeSH

Подготовка и перевод публикации
Обработка статьи · этапы, ошибки и управление
Загрузка состояния…

История · было → стало

Состояние обновляется каждые 5 секунд, пока блок открыт. Процент показывается только там, где известен объём работы. Пауза останавливает продолжение этапа; текущий запрос к источнику может завершиться.

Тип публикации указан по данным источника. Уровень доказательности автоматически не присваивается.