Coronary computed tomography angiography and endocardial leads - Image quality in 320-row CT using iterative reconstruction.

Garmer Marietta, Bonsels Marc, Metz Frauke, Klein-Wiele Oliver, Brandts Bodo, Grönemeyer Dietrich

СтатьяПеревод пока не подготовлен

Аннотация

PURPOSE: To investigate whether the presence of endocardial leads has an impact on image quality in coronary computed tomography angiography (CCTA), when current technique is employed using a 320-row computed tomography and iterative reconstruction.

MATERIALS AND

METHODS

CCTA was performed in 1641 patients, from these we identified 51 patients (study group) with endocardial leads and 51 matched partners (control group) without endocardial leads. Noise was determined in the ascending aorta and the left ventricle; signal-to-noise ratio (SNR) and contrast-to-noise ratio (CNR) were determined in the left and right coronary artery. Subjective image quality was rated separately for the 15 segments of the coronary arteries by 2 radiologists.

RESULTS

Current CCTA technique still shows slight impairment of objective image quality in patients with endocardial leads with inferior SNR in the aorta (median 15.04 versus 16.6; p = 0.004) and inferior SNR in the left/right coronary artery (median 15.3/13.81 versus 16.1/15.41; p = 0.013/0.002). CNR of the left/right coronary artery was also inferior (median 17.4/16.46 versus 19.26/19.24; p = 0.002/<0.001). The subjective image quality was rated significantly inferior only in segment 8 (p = 0.001) compared to the control group. Artifacts by ventricular leads were found in 65% of the patients in segment 8 with non-diagnostic rating in 9 cases (18%). Atrial leads resulted in artifacts predominantly in segment 1 (45%) with non-diagnostic rating in only 2 cases (4%).

CONCLUSION

CCTA is feasible with slight restrictions for patients in the presence of implanted cardiac devices when current technique is used.

Для цитирования: Garmer Marietta, Bonsels Marc, Metz Frauke, Klein-Wiele Oliver, Brandts Bodo, Grönemeyer Dietrich. Coronary computed tomography angiography and endocardial leads - Image quality in 320-row CT using iterative reconstruction.. Clinical imaging. 2018. DOI: 10.1016/j.clinimag.2018.03.006.

Предметные рубрики MeSH

Подготовка и перевод публикации
Обработка статьи · этапы, ошибки и управление
Загрузка состояния…

История · было → стало

Состояние обновляется каждые 5 секунд, пока блок открыт. Процент показывается только там, где известен объём работы. Пауза останавливает продолжение этапа; текущий запрос к источнику может завершиться.

Тип публикации указан по данным источника. Уровень доказательности автоматически не присваивается.