DOI 10.1016/j.resuscitation.2018.01.042
The validation of simplified EEG derived from the bispectral index monitor in post-cardiac arrest patients.
Аннотация
AIMS: We aimed to validate retrospectively the accuracy of simplified electroencephalography (EEG) monitoring derived from the bispectral index (BIS) monitor in post-cardiac arrest (CA) patients.
METHODS
Successfully resuscitated CA patients were transferred to the Catherization Lab followed by percutaneous coronary intervention when indicated. On arrival at the coronary care unit, bilateral BIS monitoring was started and continued up to 72 h. Raw simplified EEG tracings were extracted from the BIS monitor at a time point coinciding with the registration of standard EEG monitoring. BIS EEG tracings were reviewed by two neurophysiologists, who were asked to indicate the presence of following patterns: diffuse slowing rhythm, burst suppression pattern, cerebral inactivity, periodic epileptiform discharges and status epilepticus (SE). Additionally, these simplified BIS EEG tracings were analysed by two inexperienced investigators, who were asked to indicate the presence of SE only.
RESULTS
Thirty-two simplified BIS EEG samples were analysed. Compared to standard EEG, neurophysiologists interpreted all simplified EEG samples with a sensitivity of 86%, a specificity of 100% and an interobserver variability of 0.843. Furthermore, SE was identified with a sensitivity of 80% and a specificity of 94% by two unexperienced physicians.
CONCLUSION
Using a simple classification system, raw simplified EEG derived from a BIS monitoring device is comparable to standard EEG monitoring. Moreover, investigators without EEG experience were capable to identify SE in post-CA patients. Future studies will be warranted to confirm our results and to determine the added value of using simplified BIS EEG in terms of prognostic and therapeutic implications.
Предметные рубрики MeSH
Подготовка и перевод публикации
Обработка статьи · этапы, ошибки и управление
Ошибки и ограничения
История · было → стало
Состояние обновляется каждые 5 секунд, пока блок открыт. Процент показывается только там, где известен объём работы. Пауза останавливает продолжение этапа; текущий запрос к источнику может завершиться.
Тип публикации указан по данным источника. Уровень доказательности автоматически не присваивается.