P-wave and interatrial block: New predictor for atrial high rate episodes in patients with cardiac implantable electronic devices.

Rubio Campal José M, Benezet-Mazuecos Juan, Iglesias Bravo José Antonio, Sánchez Borque Pepa, Miracle Blanco Ángel, de la Vieja Alarcón Juan José, Martínez Mariscal Jaime, Baranchuk Adrian M, Farré Muncharaz Jerónimo

СтатьяПеревод пока не подготовлен

Аннотация

BACKGROUND

The presence of interatrial block (IAB) is associated with the development of atrial fibrillation (AF). The aim of this study was to determine whether P-wave duration and presence of IAB before the implantation of a cardiac implantable electronic device (CIED) are associated with the presence of atrial high rate episodes (AHRE), during long-term follow-up.

METHODS

380 patients (57% men; 75 ± 10 years) were included. IAB was defined according to the International Consensus Criteria. AHRE was defined as an episode of atrial rate ≥225 beats/min with a minimum duration of 5 minutes.

RESULTS

Documented paroxysmal AF before the implantation was present in 24% of the patients; 80% had hypertension and 32% structural heart disease. Mean P-wave duration was 123 ± 23 ms, and 39% of the patients had IAB (32% partial, 7% advanced). After a mean follow-up of 18 ± 12 months, 33% of the patients presented AHRE. Patients with AHRE had a P-wave duration significantly longer (130 ± 24 ms vs 119 ± 21 ms; P < 0.001) and a greater prevalence of IAB (53% vs 32%; P < 0.001). In a multivariate analysis, predictors of AHRE were: IAB (odds ratio [OR] 2.1; 95% confidence interval [CI] [1.3-3.4], P < 0.001) and previous paroxysmal AF (OR 2.6; 95% CI [1.5-4.3], P < 0.001). In patients without previous AF, the presence of IAB was also a significant predictor of AHRE (OR 3.1; 95% CI [1.8-5.5], P < 0.001).

CONCLUSIONS

IAB is a strong predictor of AHRE in patients with CIED. This finding is independent of the presence of prior paroxysmal AF.

Для цитирования: Rubio Campal José M, Benezet-Mazuecos Juan, Iglesias Bravo José Antonio, Sánchez Borque Pepa, Miracle Blanco Ángel, de la Vieja Alarcón Juan José, Martínez Mariscal Jaime, Baranchuk Adrian M, Farré Muncharaz Jerónimo. P-wave and interatrial block: New predictor for atrial high rate episodes in patients with cardiac implantable electronic devices.. Pacing and clinical electrophysiology : PACE. 2018. DOI: 10.1111/pace.13268.

Предметные рубрики MeSH

Подготовка и перевод публикации
Обработка статьи · этапы, ошибки и управление
Загрузка состояния…

История · было → стало

Состояние обновляется каждые 5 секунд, пока блок открыт. Процент показывается только там, где известен объём работы. Пауза останавливает продолжение этапа; текущий запрос к источнику может завершиться.

Тип публикации указан по данным источника. Уровень доказательности автоматически не присваивается.