Interference of heart and transcutaneous oxygen monitoring in the measurement of bioelectrical impedance analysis in preterm newborns.

Comym Viviane C, Macedu Yuri S, Neves Eduardo K P B, Bueno Arnaldo C, Fernandez Herminia C, Moreira Maria E L, Vieira Alan A

СтатьяРандомизированное исследованиеПеревод пока не подготовлен

Аннотация

OBJECTIVE

To verify if the connection of electrodes for heart and transcutaneous oxygen monitoring interfere with the measurement of electrical bioimpedance in preterm newborns.

METHODS

This was a prospective, blinded, controlled, cross-sectional, crossover study that assessed and compared paired measures of resistance (R) and reactance (Xc) by BIA, obtained with and without monitoring wires attached to the preterm newborn. The measurements were performed in immediate sequence, after randomization to the presence or absence of electrodes. The sample size calculated was 114 measurements or tests with monitoring wires and 114 without monitoring wires, considering for a difference between the averages of 0.1 ohms, with an alpha error of 10% and beta error of 20%, with significance <0.05.

RESULTS

No differences were observed between the R (677.37±196.07 vs. 677.46±194.86) and Xc (31.15±9.36 vs. 31.01±9.56) values obtained with and without monitoring wires, respectively, with good correlation between them (R: 0.997 and Xc: 0.968).

CONCLUSION

The presence of heart and/or transcutaneous oxygen monitoring wires connected to the preterm newborn did not affect the values of R or Xc measured by BIA, allowing them to be carried out in this population without risks.

Для цитирования: Comym Viviane C, Macedu Yuri S, Neves Eduardo K P B, Bueno Arnaldo C, Fernandez Herminia C, Moreira Maria E L, Vieira Alan A. Interference of heart and transcutaneous oxygen monitoring in the measurement of bioelectrical impedance analysis in preterm newborns.. Jornal de pediatria. 2016. DOI: 10.1016/j.jped.2016.01.009.

Предметные рубрики MeSH

Подготовка и перевод публикации
Обработка статьи · этапы, ошибки и управление
Загрузка состояния…

История · было → стало

Состояние обновляется каждые 5 секунд, пока блок открыт. Процент показывается только там, где известен объём работы. Пауза останавливает продолжение этапа; текущий запрос к источнику может завершиться.

Тип публикации указан по данным источника. Уровень доказательности автоматически не присваивается.