OBJECTIVE
We report the first case description of surgical aortic valve replacement after percutaneous valve implantation.
METHODS
An 87-year-old man with severe aortic stenosis who was rejected for surgical intervention underwent percutaneous valve implantation through a retrograde femoral approach. The procedure was complicated by cardiogenic shock caused by severe aortic insufficiency, leading to emergency surgical aortic valve replacement.
RESULTS
The operative findings revealed the presence of commissural paravalvular leaks and centrally malapposed leaflets. Surgical replacement was uneventful, and the patient was discharged on day 30, despite a challenging postoperative course. His follow-up at 1 year has been uneventful. This case illustrates that overdilatation of the stent is not recommended because it might worsen central aortic insufficiency. Moreover, the transapical route should be considered when the appropriately sized prosthesis is unable to be inserted because of inappropriate vascular access. However, despite an initial "prohibitive" surgical risk, surgical aortic valvular replacement after percutaneous valve implantation could be easily performed.
CONCLUSION
Percutaneous heart valve implantation, which provides a larger surface area than balloon valvotomy, can be offered to patients with cardiogenic shock and severe comorbidities to improve their hemodynamic state and reduce their surgical risk.
Предметные рубрики MeSH
Подготовка и перевод публикации
Обработка статьи · этапы, ошибки и управление
Ошибки и ограничения
История · было → стало
Состояние обновляется каждые 5 секунд, пока блок открыт. Процент показывается только там, где известен объём работы. Пауза останавливает продолжение этапа; текущий запрос к источнику может завершиться.
Тип публикации указан по данным источника. Уровень доказательности автоматически не присваивается.