Procoagulant properties of flow fields in stenotic and expansive orifices.

Fallon Anna M, Dasi Lakshmi Prasad, Marzec Ulla M, Hanson Stephen R, Yoganathan Ajit P

СтатьяПоддержка исследования: вне правительства СШАПеревод пока не подготовлен

Аннотация

In the United States, over 125,000 mechanical heart valves (MHVs) are implanted each year. Flow through the MHV hinge can cause thromboemboli formation. The purpose of this study was to examine various orifice geometries representing the MHV hinge region and how these geometries may contribute to platelet activation and thrombin generation. We also characterized these flow fields with digital particle image velocimetry (DPIV). Citrated human blood at room temperature was forced through the orifices (400 and 800 microm ID) with a centrifugal bypass pump, continuously infusing calcium chloride to partially reverse the citrate anticoagulant. Blood samples were tested for the presence of thrombin-antithrombin complex (TAT) and platelet factor 4 (PF4). Velocity and shear stress were measured with DPIV using a blood analog fluid seeded with fluorescent microbeads. The results indicate that small changes in geometry, although they do not affect the bulk flow, change the coagulation propensity as blood flows through the orifices. A more abrupt geometry allows more stagnation to occur resulting in more thrombin generation. PF4 measurements indicated similar levels of platelet activation for all orifices. DPIV showed differences in the jets with respect to entrainment of stagnant fluid. These results help to pinpoint the important parameters that lead to flow stasis and subsequent thrombus formation.

Для цитирования: Fallon Anna M, Dasi Lakshmi Prasad, Marzec Ulla M, Hanson Stephen R, Yoganathan Ajit P. Procoagulant properties of flow fields in stenotic and expansive orifices.. Annals of biomedical engineering. 2007. DOI: 10.1007/s10439-007-9398-3.

Предметные рубрики MeSH

Подготовка и перевод публикации
Обработка статьи · этапы, ошибки и управление
Загрузка состояния…

История · было → стало

Состояние обновляется каждые 5 секунд, пока блок открыт. Процент показывается только там, где известен объём работы. Пауза останавливает продолжение этапа; текущий запрос к источнику может завершиться.

Тип публикации указан по данным источника. Уровень доказательности автоматически не присваивается.