DOI 10.1159/000503924

Медицинская библиотека · Kidney & blood pressure research · 2019

Monitoring of Intraperitoneal Fluid Volume during Peritoneal Equilibration Testing using Segmental Bioimpedance.

Zhu Fansan, Abbas Samer R, Bologa Roxana M, Levin Nathan W, Kotanko Peter

СтатьяПеревод пока не подготовлен

Аннотация

BACKGROUND

Ultrafiltration failure and fluid overload are common in peritoneal dialysis (PD) patients. Knowledge of intraperitoneal volume (IPV) and time to peak IPV during a dwell would permit improved PD prescription. This study aimed to utilize trunk segmental bioimpedance analysis (SBIA) to quasi-continuously monitor IPV (IPVSBIA) during the peritoneal dwell.

METHODS

IPVSBIA was measured every minute using lower-trunk SBIA (Hydra 4200; Xitron Technologies Inc., CA, USA) in 10 PD patients during a standard 240-min peritoneal equilibration test (PET). The known dialysate volume (2 L) rendered IPVSBIA calibration and calculation of instantaneous ultrafiltration volume (UFVSBIA) possible. UFVSBIA was defined as IPVSBIA - 2 L.

RESULTS

Based on dialysate-to-plasma creatinine ratio, 2 patients were high, 7 high-average, and 1 low-average transporters. Technically sound IPVSBIA measurements were obtained in 9 patients (age 59.0 ± 8.8 years, 7 females, 5 African Americans). Drained ultrafiltration volume (UFVdrain) was 0.47 ± 0.21 L and correlated (r = 0.74; p < 0.05) with end-dwell UFVSBIA (0.55 ± 0.17 L). Peak UFVSBIA was 1.04 ± 0.32 L, it was reached 177 ± 61 min into the dwell and exceeded end-dwell UFVSBIA by 0.49 ± 0.28 L (95% CI: 0.27-0.7) and UFVdrain by 0.52 ± 0.31 L (95% CI: 0.29-0.76), respectively.

CONCLUSION

This pilot study demonstrates the feasibility of trunk segmental bioimpedance to quasi-continuously monitor IPVSBIA and identify the time to peak UFVSBIA during a standard PET. Such new insights into the dynamics of intraperitoneal fluid volume during the dwell may advance our understanding of the underlying transport physiology and eventually assist in improving PD treatment prescriptions.

Для цитирования: Zhu Fansan, Abbas Samer R, Bologa Roxana M, Levin Nathan W, Kotanko Peter. Monitoring of Intraperitoneal Fluid Volume during Peritoneal Equilibration Testing using Segmental Bioimpedance.. Kidney & blood pressure research. 2019. DOI: 10.1159/000503924.

Предметные рубрики MeSH

Подготовка и перевод публикации
Обработка статьи · этапы, ошибки и управление
Загрузка состояния…

История · было → стало

Состояние обновляется каждые 5 секунд, пока блок открыт. Процент показывается только там, где известен объём работы. Пауза останавливает продолжение этапа; текущий запрос к источнику может завершиться.

Тип публикации указан по данным источника. Уровень доказательности автоматически не присваивается.