Multicenter study of the safety and effects of magnetic resonance imaging in patients with coronary sinus left ventricular pacing leads.

Sheldon Seth H, Bunch T Jared, Cogert Gregory A, Acker Nancy G, Dalzell Connie M, Higgins John V, Espinosa Raul E, Asirvatham Samuel J, Cha Yong-Mei, Felmlee Joel P, Watson Robert E, Anderson Jeffrey L, Brooks Miriam H, Osborn Jeffrey S, Friedman Paul A

СтатьяМногоцентровое исследованиеПеревод пока не подготовлен

Аннотация

BACKGROUND

Magnetic resonance imaging (MRI) in patients with left ventricular (LV) leads may cause tissue or lead heating, dislodgment, venous damage, or lead dysfunction.

OBJECTIVE

The purpose of this study was to determine the safety of MRI in patients with LV pacing leads.

METHODS

Prospective data on patients with coronary sinus LV leads undergoing clinically indicated MRI at 3 institutions were collected. Patients were not pacemaker-dependent. Scans were performed under pacing nurse, technician, radiologist, and physicist supervision using continuous vital sign, pulse oximetry, and ECG monitoring and a 1.5-T scanner with specific absorption rate <1.5 W/kg. Devices were interrogated pre- and post-MRI, programmed to asynchronous or inhibition mode with tachyarrhythmia therapies off (if present), and reprogrammed to their original settings post-MRI.

RESULTS

MRI scans (n = 42) were performed in 40 patients with non-MRI conditional LV leads between 2005 and 2013 (mean age 67 ± 9 years, n = 16 [40%] women, median lead implant duration 740 days with interquartile range 125-1173 days). MRIs were performed on the head/neck/spine (n = 35 [83%]), lower extremities (n = 4 [10%]), chest (n = 2 [5%]), and abdomen (n = 1 [2%]). There were no overall differences in pre- and post-MRI interrogation LV lead sensing (12.4 ± 6.2 mV vs 12.9 ± 6.7 mV, P = .38), impedance (724 ± 294 Ω vs 718 ± 312 Ω, P = .67), or threshold (1.4 ± 1.1 V vs 1.4 ± 1.0 V, P = .91). No individual LV lead changes required intervention.

CONCLUSION

MRI scanning was performed safely in non-pacemaker-dependent patients with coronary sinus LV leads who were carefully monitored during imaging without clinically significant adverse effect on LV lead function.

Для цитирования: Sheldon Seth H, Bunch T Jared, Cogert Gregory A, Acker Nancy G, Dalzell Connie M, Higgins John V, Espinosa Raul E, Asirvatham Samuel J, Cha Yong-Mei, Felmlee Joel P, Watson Robert E, Anderson Jeffrey L, Brooks Miriam H, Osborn Jeffrey S, Friedman Paul A. Multicenter study of the safety and effects of magnetic resonance imaging in patients with coronary sinus left ventricular pacing leads.. Heart rhythm. 2014. DOI: 10.1016/j.hrthm.2014.11.037.

Предметные рубрики MeSH

Подготовка и перевод публикации
Обработка статьи · этапы, ошибки и управление
Загрузка состояния…

История · было → стало

Состояние обновляется каждые 5 секунд, пока блок открыт. Процент показывается только там, где известен объём работы. Пауза останавливает продолжение этапа; текущий запрос к источнику может завершиться.

Тип публикации указан по данным источника. Уровень доказательности автоматически не присваивается.