Motion mitigation in positron-emission tomography guided radiotherapy delivered on a magnetic resonance imaging-linear accelerator.

Everard Albert J, Santo Rodrigo José, Borman Pim T S, Uijtewaal Prescilla, Fast Martin F, de Jong Hugo W A M, van den Berg Cornelis A T, Raaymakers Bas W, Beijst Casper

СтатьяПеревод пока не подготовлен

Аннотация

BACKGROUND AND PURPOSE: The precision of Positron Emission Tomography (PET)-guided radiotherapy may be compromised by motion. To address this, we demonstrated the feasibility of a motion-robust PET-to-magnetic resonance linear accelerator (MR-linac) pipeline.

MATERIALS AND

METHODS

For comparison, PET images of a phantom with a target in a movable insert were acquired under a patient-derived breathing waveform and static conditions. Motion correction using the PET Maximum-Likelihood Motion and Activity (MLMA) reconstruction was compared with static reference and motion-blurred images. Targets were delineated using thresholding based on maximum standardized uptake value, and Intensity-Modulated Radiation Therapy (IMRT) plans were generated. Plans were delivered on an MR-linac with and without gating or Multi Leaf Collimator (MLC) tracking. Dosimetry was assessed using film measurements, evaluating dose differences and 3%/3 mm gamma pass rates. The pipeline was further evaluated using clinical PET data, delivered with and without gating and tracking after motion correction. Dosimetric evaluation was performed using a cylindrical multi-diode phantom.

RESULTS

Motion-blurred PET produced larger gross target volumes (GTV) (GTV: 58 cm3, GTVboost: 14 cm3) than motion-corrected PET (GTV: 52 cm3, GTVboost: 8 cm3), the latter matching the static reference (GTV: 52 cm3, GTVboost: 9 cm3). Motion correction reduced dose discrepancies and improved gamma pass rates (72%), with further gains from gating (90.3%) and MLC tracking (98.3%). The measurements of patient simulations showed higher gamma pass rates for the motion-corrected plan (static-delivery: 81.8%, gating: 99.2%, MLC-tracked: 100%) compared to motion-blurred plan (static-delivery: 49.4%, gating: 65.6%, MLC-tracked: 63.6%).

CONCLUSIONS

A motion-robust PET-to-MR-linac pipeline using motion-corrected PET imaging and MR-guided motion mitigation is technically feasible and enhances dose precision.

Для цитирования: Everard Albert J, Santo Rodrigo José, Borman Pim T S, Uijtewaal Prescilla, Fast Martin F, de Jong Hugo W A M, van den Berg Cornelis A T, Raaymakers Bas W, Beijst Casper. Motion mitigation in positron-emission tomography guided radiotherapy delivered on a magnetic resonance imaging-linear accelerator.. Physics and imaging in radiation oncology. 2026. DOI: 10.1016/j.phro.2026.100970.
Подготовка и перевод публикации
Обработка статьи · этапы, ошибки и управление
Загрузка состояния…

История · было → стало

Состояние обновляется каждые 5 секунд, пока блок открыт. Процент показывается только там, где известен объём работы. Пауза останавливает продолжение этапа; текущий запрос к источнику может завершиться.

Тип публикации указан по данным источника. Уровень доказательности автоматически не присваивается.