MedElevator Войти в исследования
← Поиск публикаций

DOI 10.1007/s10047-013-0691-7

Медицинская библиотека · Journal of artificial organs : the official journal of the Japanese Society for Artificial Organs · 2013

Continuous-flow pump model study: the effect on pump performance of pump characteristics and cardiovascular conditions.

Ferrari Gianfranco, Kozarski Maciej, Fresiello Libera, Di Molfetta Arianna, Zieliński Krzysztof, Górczyńska Krystyna, Pałko Krzysztof J, Darowski Marek

СтатьяПоддержка исследования: вне правительства СШАПеревод пока не подготовлен

Аннотация

This model study evaluates the effect of pump characteristics and cardiovascular data on hemodynamics in atrio-aortic VAD assistance. The model includes a computational circulatory sub-model and an electrical sub-model representing two rotary blood pumps through their pressure-flow characteristics. The first is close to a pressure generator-PG (average flow sensitivity to pressure variations, -0.047 l mmHg(-1)); the second is closer to a flow generator-FG (average flow sensitivity to pressure variations, -0.0097 l mmHg(-1)). Interaction with VAD was achieved by means of two interfaces, behaving as impedance transformers. The model was verified by use of literature data and VAD onset conditions were used as a control for the experiments. Tests compared the two pumps, at constant pump speed, in different ventricular and circulatory conditions: maximum ventricular elastance (0.44-0.9 mmHg cm(-3)), systemic peripheral resistance (781-1200 g cm(-4) s(-1)), ventricular diastolic compliance C p (5-10-50 cm(3) mmHg(-1)), systemic arterial compliance (0.9-1.8 cm(3) mmHg(-1)). Analyzed variables were: arterial and venous pressures, flows, ventricular volume, external work, and surplus hemodynamic energy (SHE). The PG pump generated the highest SHE under almost all conditions, in particular for higher C p (+50 %). PG pump flow is also the most sensitive to E max and C p changes (-26 and -33 %, respectively). The FG pump generally guarantees higher external work reduction (54 %) and flow less dependent on circulatory and ventricular conditions. The results are evidence of the importance of pump speed regulation with changing ventricular conditions. The computational sub-model will be part of a hydro-numerical model, including autonomic controls, designed to test different VADs.

Для цитирования: Ferrari Gianfranco, Kozarski Maciej, Fresiello Libera, Di Molfetta Arianna, Zieliński Krzysztof, Górczyńska Krystyna, Pałko Krzysztof J, Darowski Marek. Continuous-flow pump model study: the effect on pump performance of pump characteristics and cardiovascular conditions.. Journal of artificial organs : the official journal of the Japanese Society for Artificial Organs. 2013. DOI: 10.1007/s10047-013-0691-7.

Предметные рубрики MeSH

Подготовка и перевод публикации
Обработка статьи · этапы, ошибки и управление
Загрузка состояния…

История · было → стало

Состояние обновляется каждые 5 секунд, пока блок открыт. Процент показывается только там, где известен объём работы. Пауза останавливает продолжение этапа; текущий запрос к источнику может завершиться.

Тип публикации указан по данным источника. Уровень доказательности автоматически не присваивается.