Nonphysiologic noise early after defibrillator implantation in Canada: incidence and implications: a report from the Canadian Heart Rhythm Society Device Committee.

Essebag Vidal, Champagne Jean, Birnie David H, Verma Atul, Healey Jeffrey S, Simpson Christopher S, Kus Teresa, Thibault Bernard, Mangat Iqwal, Tung Stanley, Sterns Laurence, Exner Derek V, Davies Ted, Coutu Benoit, Crystal Eugene, Stephenson Elizabeth A, Connors Sean, Paredes Felix Ayala, Parkash Ratika, Krahn Andrew D

Клинический случайСтатьяПоддержка исследования: вне правительства СШАПеревод пока не подготовлен

Аннотация

BACKGROUND

Following recent cases of nonphysiologic noise noted early after defibrillator implantation, the Canadian Heart Rhythm Society Device Committee decided to evaluate the implications of this problem.

OBJECTIVE

To determine the incidence and consequences of nonphysiologic noise early after defibrillator implantation.

METHODS

The Canadian Heart Rhythm Society Device Committee surveyed all Canadian defibrillator implanting centers regarding their implant volumes and number of cases where nonphysiologic noise had been noted early (< 24 hours) after implant over the preceding 2 years. For such cases, information regarding the manufacturer and occurrence of inappropriate shock or inhibition of pacing was reported.

RESULTS

Responses were obtained from 20 of 23 surveyed implanting centers, with a total implant volume of 4960 defibrillators per year. The occurrence of nonphysiologic noise early after implantation was noted in 25 cases over the preceding 2 years (0.25%). Noise was detected in devices of all 3 of the leading volume device manufacturers. There were 2 cases of inappropriate shocks and 2 cases of symptomatic pacing inhibition. In 4 cases, removal of the lead from the header and retesting with the analyzer confirmed normal lead function. In all cases, the noise resolved within 24 hours.

CONCLUSION

Nonphysiologic noise early after defibrillator implantation was noted in 0.25% of procedures and was not limited to one specific manufacturer. This noise may result in an inappropriate shock or inhibition of pacing in a pacemaker-dependent patient. This transient phenomenon, possibly related to fluid and/or air in the header, appears to always resolve without surgical intervention.

Для цитирования: Essebag Vidal, Champagne Jean, Birnie David H, Verma Atul, Healey Jeffrey S, Simpson Christopher S, Kus Teresa, Thibault Bernard, Mangat Iqwal, Tung Stanley, Sterns Laurence, Exner Derek V, Davies Ted, Coutu Benoit, Crystal Eugene, Stephenson Elizabeth A, Connors Sean, Paredes Felix Ayala, Parkash Ratika, Krahn Andrew D. Nonphysiologic noise early after defibrillator implantation in Canada: incidence and implications: a report from the Canadian Heart Rhythm Society Device Committee.. Heart rhythm. 2011. DOI: 10.1016/j.hrthm.2011.10.025.

Предметные рубрики MeSH

Подготовка и перевод публикации
Обработка статьи · этапы, ошибки и управление
Загрузка состояния…

История · было → стало

Состояние обновляется каждые 5 секунд, пока блок открыт. Процент показывается только там, где известен объём работы. Пауза останавливает продолжение этапа; текущий запрос к источнику может завершиться.

Тип публикации указан по данным источника. Уровень доказательности автоматически не присваивается.