Pacing and clinical electrophysiology : PACE · 2011

Effect of insulation material in aging pacing leads: comparison of impedance and other electricals: time-dependent pacemaker insulation changes.

СтатьяПоддержка исследования: вне правительства СШАНазвание и аннотация: Перевод пока не подготовлен
Обработка статьи · этапы, ошибки и управление
Загрузка состояния…

История · было → стало

Состояние обновляется каждые 5 секунд, пока блок открыт. Процент показывается только там, где известен объём работы. Пауза останавливает продолжение этапа; текущий запрос к источнику может завершиться.

Johnson W Ben, Braly Alan, Cobian Kenneth, Craig Mary B, Voegtlin Loline, Haddad Tarek, McVenes Rick

Аннотация

BACKGROUND: There has been concern over declining bipolar (BP) impedance (Z) in aging polyurethane (PU) cardiac pacing leads. Subsequently, a prospective study was conducted comparing BP Z, threshold (Th), and R-wave sensing amplitude of 55D PU-insulated (Model 4024, Medtronic, Inc., Minneapolis, MN, USA) and silicone-insulated (Model 5024) leads.

METHODS: This study was initiated by The Iowa Heart Center. Patients with Model 4024 (N = 162) or 5024 (N = 120) pacing leads with at least 6 years implant time were enrolled and followed for an additional 5 years.

RESULTS: There was a significant drop in the mean BP Z for the Model 4024 population, between enrollment (6 years) and the final endpoint (11 years), which was in contrast to the Model 5024 which did not see a significant drop in its mean BP Z for this same period. The trend difference seen in the means between the two models was statistically significant (P < 0.0001). In addition, a statistically significant relationship was found between dropping BP Z and rising Th (P < 0.0001). The analysis showed that if BP Z dropped below 200 ohms, the probability of having a >3X increase over baseline, in Th at 2.5 V, increases from approximately 3-7% to as high as 30%.

CONCLUSIONS: A significant drop in BP Z observed in the PU-insulated Model 4024 lead was not present in the silicone-insulated Model 5024 lead. The statistically significant relationship between dropping BP Z and rising Th helps to understand how to better manage patients with aging leads.