Determinants of Greater Peak radiation skin dose in percutaneous coronary intervention for chronic total occlusion.

Yamamoto Kei, Sakakura Kenichi, Hamamoto Kohei, Hasegawa Hiroko, Tsukui Takunori, Seguchi Masaru, Taniguchi Yousuke, Wada Hiroshi, Momomura Shin-Ichi, Fujita Hideo

СтатьяПеревод пока не подготовлен

Аннотация

BACKGROUND

Peak skin dose (PSD) is closely associated with skin radiation injuries such as skin ulcers in percutaneous coronary intervention (PCI). Although PSD is greater in PCI for chronic total occlusion (CTO) lesions as compared with non-CTO lesions, the determinants of PSD in CTO-PCI are not fully understood. The purpose of this study was to investigate the clinical factors associated with excess PSD in PCI for CTO.

METHODS

The study population included a total of 220 CTO-PCI cases that were divided into a standard PSD group (<2 Gy, n = 187) and an excess PSD group (≥2 Gy, n = 33). Clinical, lesion, and procedural characters were compared between the 2 groups. Multivariate logistic regression was performed to investigate the clinical factors associated with excess PSD.

RESULTS

Body surface area (BSA) was significantly higher in the excess PSD group (1.85 ± 0.24 m2) than the standard PSD group (1.71 ± 0.18 m2) (p = 0.001). The J-CTO score was significantly higher in the excess PSD group (2.51 ± 1.28) than the standard PSD group (1.60 ± 1.13) (p < 0.001). Multivariate logistic regression analysis revealed that BSA (0.1 mm increase: OR 1.663, 95% CI 1.300-2.128, p < 0.001) and J-CTO score (1-point increase: OR 2.015, 95% CI 1.322-3.071, p = 0.001) were significantly associated with excess PSD.

CONCLUSIONS

A large BSA and high J-CTO score were significantly associated with excess PSD. It is important to pay special attention to CTO patients who have a large BSA and/or high J-CTO score to reduce patient's PSD.

Для цитирования: Yamamoto Kei, Sakakura Kenichi, Hamamoto Kohei, Hasegawa Hiroko, Tsukui Takunori, Seguchi Masaru, Taniguchi Yousuke, Wada Hiroshi, Momomura Shin-Ichi, Fujita Hideo. Determinants of Greater Peak radiation skin dose in percutaneous coronary intervention for chronic total occlusion.. Journal of cardiology. 2020. DOI: 10.1016/j.jjcc.2020.02.021.

Предметные рубрики MeSH

Подготовка и перевод публикации
Обработка статьи · этапы, ошибки и управление
Загрузка состояния…

История · было → стало

Состояние обновляется каждые 5 секунд, пока блок открыт. Процент показывается только там, где известен объём работы. Пауза останавливает продолжение этапа; текущий запрос к источнику может завершиться.

Тип публикации указан по данным источника. Уровень доказательности автоматически не присваивается.