Upper limit of vulnerability during defibrillator implantations predicts the occurrence of appropriate shock therapy for ventricular fibrillation.

Yamashita Soichiro, Yoshida Akihiro, Fukuzawa Koji, Fujiwara Ryudo, Suzuki Atsushi, Nakanishi Tomoyuki, Matsumoto Akinori, Konishi Hiroki, Ichibori Hirotoshi, Hirata Ken-ichi

СтатьяНаблюдательное исследованиеПеревод пока не подготовлен

Аннотация

BACKGROUND

The utility of the upper limit of vulnerability (ULV) test in patients undergoing defibrillator implantation has been reported, so the purpose of this study was to evaluate the difference in the clinical outcomes between patients with ULV ≤15 J or >15 J.

METHODS AND

RESULTS

A total of 165 patients receiving an implantable cardioverter-defibrillator underwent a vulnerability test. At the time of the implantation, we delivered a 15-J shock on the T-peak and ±20 ms later to cover the most vulnerable part of the cardiac cycle. The clinical outcomes were prospectively analyzed. A 15-J shock induced ventricular fibrillation (VF) in 30 patients (ULV >15 J) and did not in 135 (ULV ≤15 J). The characteristics of the 2 groups were comparable. After a mean follow-up of 757 days, Kaplan-Meier curve analysis showed that the ULV ≤15 J group experienced less VF than the ULV >15 J group (log-rank P=0.003). The occurrence of ventricular tachycardia was similar between the 2 groups (P=0.140). Furthermore, the effectiveness of ATP was comparable. After adjusting for other known predictors of shock therapy, a ULV >15 J was independently associated with the occurrence of VF (hazard ratio: 6.25; 95% confidence interval: 1.913-20.40; P<0.01).

CONCLUSIONS

A high ULV value was associated with a high incidence of VF, which suggests that cardiac vulnerability to electrical shock may be linked to electrical instability.

Для цитирования: Yamashita Soichiro, Yoshida Akihiro, Fukuzawa Koji, Fujiwara Ryudo, Suzuki Atsushi, Nakanishi Tomoyuki, Matsumoto Akinori, Konishi Hiroki, Ichibori Hirotoshi, Hirata Ken-ichi. Upper limit of vulnerability during defibrillator implantations predicts the occurrence of appropriate shock therapy for ventricular fibrillation.. Circulation journal : official journal of the Japanese Circulation Society. 2014. DOI: 10.1253/circj.cj-14-0136.

Предметные рубрики MeSH

Подготовка и перевод публикации
Обработка статьи · этапы, ошибки и управление
Загрузка состояния…

История · было → стало

Состояние обновляется каждые 5 секунд, пока блок открыт. Процент показывается только там, где известен объём работы. Пауза останавливает продолжение этапа; текущий запрос к источнику может завершиться.

Тип публикации указан по данным источника. Уровень доказательности автоматически не присваивается.