Radiation tolerance of contemporary implantable cardioverter-defibrillators.

Mollerus Michael, Naslund Leslee, Lipinski Margaret, Meyer Anne, Libey Bruce, Dornfeld Ken

Сравнительное исследованиеОценочное исследованиеСтатьяПоддержка исследования: вне правительства СШАПеревод пока не подготовлен

Аннотация

BACKGROUND

Implantable cardioverter-defibrillators (ICDs) are complex instruments using integrated circuit technology. Previous studies suggested risk to the device when exposed to a radiation environment. Little data is available on contemporary ICD systems.

OBJECTIVES

The purpose of the present study was to assess the ability of contemporary ICD designs to resist the damaging effects of direct exposure to therapeutic doses of radiation.

METHODS

Four contemporary ICDs and four legacy ICDs devices were exposed to escalating doses of photon irradiation (XRT) from a 6-MV linear accelerator. Escalating doses were administered over 8 days to a maximum cumulative dose of 131.11 Gy or catastrophic failure.

RESULTS

Each legacy device had catastrophic failure following the 6th XRT session, characterized by failure to deliver shock therapy. All four contemporary devices remained fully functional following the 8th and final XRT session (P = 0.03). The cumulative, survived radiation dose was significantly different between the contemporary and legacy groups (131.11 vs. 41.11 Gy, P = 0.01). Changes seen in the legacy devices were sudden and not anticipated by trends in prior sessions.

CONCLUSION

The results of this study suggest that contemporary ICD designs may be more robust than earlier designs in a radiation environment.

Для цитирования: Mollerus Michael, Naslund Leslee, Lipinski Margaret, Meyer Anne, Libey Bruce, Dornfeld Ken. Radiation tolerance of contemporary implantable cardioverter-defibrillators.. Journal of interventional cardiac electrophysiology : an international journal of arrhythmias and pacing. 2013. DOI: 10.1007/s10840-013-9861-z.

Предметные рубрики MeSH

Подготовка и перевод публикации
Обработка статьи · этапы, ошибки и управление
Загрузка состояния…

История · было → стало

Состояние обновляется каждые 5 секунд, пока блок открыт. Процент показывается только там, где известен объём работы. Пауза останавливает продолжение этапа; текущий запрос к источнику может завершиться.

Тип публикации указан по данным источника. Уровень доказательности автоматически не присваивается.