Effects of high-dose radiotherapy on implantable cardioverter defibrillators: an in vivo porcine study.

Zaremba Tomas, Jakobsen Annette R, Thøgersen Anna M, Riahi Sam, Kjærgaard Benedict

СтатьяПоддержка исследования: вне правительства СШАПеревод пока не подготовлен

Аннотация

INTRODUCTION

Although the effects of radiotherapy (RT) on implantable cardioverter defibrillators (ICDs) have been studied in vitro, and some information has been gathered from case reports and series, no in vivo experiments have previously been performed.

METHODS

In vivo effects of photon RT applied directly to five modern ICD generators from different manufacturers implanted in pigs were studied. The devices were interrogated between and after increasing doses of ionizing radiation. Afterwards, the shock function was tested.

RESULTS

All ICDs withstood fractionated irradiations with a cumulative dose of 18.5 gray (Gy) of 6 megavolt (MV) photons and 18.5 Gy of 18 MV photons and were still fully functional. Especially, no oversense was recorded. Induced ventricular fibrillation was detected and treated properly by shock therapy in all cases. However, one of the ICDs converted to back-up mode later the same night.

CONCLUSION

The animal model is feasible for investigating RT effects on implanted cardiac devices. During irradiations with 37 Gy, one recoverable malfunction was present in the tested devices. Additional animal studies could provide supplementary evidence for treating ICD patients, including recommendations for reprogramming of the ICD during RT and avoidance of relocating the device.

Для цитирования: Zaremba Tomas, Jakobsen Annette R, Thøgersen Anna M, Riahi Sam, Kjærgaard Benedict. Effects of high-dose radiotherapy on implantable cardioverter defibrillators: an in vivo porcine study.. Pacing and clinical electrophysiology : PACE. 2013. DOI: 10.1111/pace.12249.

Предметные рубрики MeSH

Подготовка и перевод публикации
Обработка статьи · этапы, ошибки и управление
Загрузка состояния…

История · было → стало

Состояние обновляется каждые 5 секунд, пока блок открыт. Процент показывается только там, где известен объём работы. Пауза останавливает продолжение этапа; текущий запрос к источнику может завершиться.

Тип публикации указан по данным источника. Уровень доказательности автоматически не присваивается.