DOI 10.1097/j.jcrs.0000000000000910
Clinical and refractive outcomes after topography-guided refractive surgery planned using Phorcides surgery planning software.
Аннотация
PURPOSE: To evaluate prospectively the clinical outcomes of topography-guided laser in situ keratomileusis (LASIK) surgery performed with Contoura Vision using the Wavelight excimer laser and planned with the Phorcides Analytical Engine.
SETTING
Four clinical practices in the U.S.
DESIGN
Prospective single-arm interventional study.
METHODS
130 eyes of 65 patients with myopia or myopic astigmatism were enrolled in a prospective study of visual and refractive outcomes after treatment with Contoura Vision using the Wavelight EX500 excimer laser to achieve optimal distance vision. At 3 months postoperatively, uncorrected distance visual acuity (UDVA), manifest refraction, and corrected distance visual acuity (CDVA) were measured and compared with previously published studies.
RESULTS
At 3 months postoperatively, 100%, 89%, and 28% of eyes achieved UDVA of 20/20, 20/15, and 20/12.5 or better, respectively. 92% of eyes had postoperative UDVA equal to or better than their preoperative CDVA. Postoperative CDVA was equal to, 1 line better, or 2 lines better than the preoperative CDVA in 53%, 40%, and 6% of eyes, respectively. Only 1 eye lost 1 line of CDVA, and no eyes lost more than 1 line of CDVA. A statistically significant decrease was noted in subjective visual complaints, including glare, halos, difficulty driving at night, reading difficulty, starbursts, fluctuation in vision, and light sensitivity.
CONCLUSIONS
The Phorcides Analytical Engine can be used to optimize visual outcomes for the correction of myopia and myopic astigmatism.