DOI 10.1016/j.mjafi.2020.03.014
Rotational stability of Toric intraocular lenses.
Аннотация
BACKGROUND
A prospective study to evaluate the rotational stability of toric intraocular lenses (IOLs).
METHODS
A prospective study of 30 eyes in 29 patients. All patients with regular astigmatism of range 0.75-4 D were included in the study. Exclusion criteria included irregular corneal astigmatism, post-refractive surgery and corneal dystrophies. All patients underwent uncomplicated phacoemulsification cataract surgery by the same surgeon. Three eyes with corneal astigmatism of 0.75-1.5 D were implanted with the AcrySof SA60T3 IOL, eight eyes with astigmatism between 1.5 and 2.0 D received the SA60T4 IOL and eight eyes with astigmatism between 2.0 and 2.50 D received the SA60T5 and rest with astigmatism of 2.5 D and higher received SA60T6 and above models. Main outcome measure was the post-operative position of the lens, assessed at day 1, 1 week, 1 month and 6 months, using toric marker and the slit lamp.
RESULTS
There was no significant rotation of IOL observed during a follow-up period of 6 months. Overall, the post-operative rotation was within 5° in 95% of cases. There was no trend for either clockwise or anti-clockwise rotation. However, IOL rotation happens mostly within the first month of surgery, and if it is significant, it requires early repositioning.
CONCLUSION
Toric IOLs are very effective and consistent in correcting astigmatism during the cataract surgery. Once placed to its position, toric IOLs demonstrate rotational stability in the capsular bag.
Подготовка и перевод публикации
Обработка статьи · этапы, ошибки и управление
Ошибки и ограничения
История · было → стало
Состояние обновляется каждые 5 секунд, пока блок открыт. Процент показывается только там, где известен объём работы. Пауза останавливает продолжение этапа; текущий запрос к источнику может завершиться.
Тип публикации указан по данным источника. Уровень доказательности автоматически не присваивается.