DOI 10.1097/j.jcrs.0000000000000819

Медицинская библиотека · Journal of cataract and refractive surgery · 2021

Predicted vs measured posterior corneal astigmatism for toric intraocular lens calculations.

Shammas H John, Yu Fei, Shammas Maya C, Jivrajka Renu, Hakimeh Christine

СтатьяНаблюдательное исследованиеПеревод пока не подготовлен

Аннотация

PURPOSE: To evaluate the astigmatic correction obtained with a toric intraocular lens using the keratometric readings (Ks) from a swept-source optical coherence tomography (SS-OCT) biometer and the Barrett toric formula with its predicted posterior corneal astigmatism (PCA) value and to compare the results with those expected by using the OCT Ks and a measured PCA from a scheimpflug topographer and by using the SimKs and the measured PCA from the Scheimpflug topographer.

SETTING

Private practice, Lynwood, California.

DESIGN

Retrospective observational study.

METHODS

All measurements were performed by the SS-OCT biometer and the Scheimpflug topographer and using the Barrett toric formula.

RESULTS

We evaluated 122 eyes of 122 patients. The mean absolute errors in predicted residual astigmatism for the entire series were 0.41 ± 0.19 diopters (D) (0.00 to 0.85 D) using the OCT Ks and predicted PCA, 0.45 ± 0.25 D (0.00 to 1.01 D) using the OCT Ks and measured PCA, and 0.49 ± 0.25 D (0.00 to 1.30 D) using the SimKs and measured PCA. The statistically significant differences between the errors had a P value of .062 for the entire series (n = 122), .26 for the subgroup with against-the-rule astigmatism (n = 68), .47 for the subgroup with oblique astigmatism (n = 11), and .05 for the subgroup with with-the-rule astigmatism (n = 43). The percentage of eyes within ±0.50 D were 74% (n = 90), 71% (n = 87) and 64% (n = 78) (P = .13) and within ±0.75 D were 99% (n = 121), 95% (n = 116) and 84% (n = 102) (P < .001), respectively.

CONCLUSIONS

The Barrett toric formula and its predicted PCA performed better with the OCT K readings than with the topographer SimKs and a measured PCA.

Для цитирования: Shammas H John, Yu Fei, Shammas Maya C, Jivrajka Renu, Hakimeh Christine. Predicted vs measured posterior corneal astigmatism for toric intraocular lens calculations.. Journal of cataract and refractive surgery. 2021. DOI: 10.1097/j.jcrs.0000000000000819.