The Precision of a New Anterior Segment Optical Coherence Tomographer and Its Comparison With a Swept-Source OCT-Based Optical Biometer in Patients With Cataract.

Wang Qinmei, Chen Min, Ning Rui, Savini Giacomo, Wang Yiran, Zhang Tianjiao, Lin Xuanqiao, Chen Xin, Huang Jinhai

СтатьяПеревод пока не подготовлен

Аннотация

PURPOSE: To assess the precision of a new spectral-domain optical coherence tomographer (SD-OCT)/Placido topographer (MS-39; CSO) and its comparison with a swept-source OCT (SS-OCT) biometer (Argos; Movu, Inc) in patients with cataract.

METHODS

Fifty-three right eyes from 53 patients were examined by two experienced operators three times using both devices randomly. Employing the within-subject standard deviation (Sw), test-retest variability, coefficient of variation, and intraclass correlation coefficient to evaluate intraoperator repeatability and interoperator reproducibility; the double-angle plots to analyze astigmatism; and Bland-Altman plots and 95% limits of agreement to verify the agreement between devices.

RESULTS

The SD-OCT/Placido tomographer showed high precision, with coefficient of variation of 0.44% or less, intraclass correlation coefficient of 0.945 or greater for all parameters, test-retest variability of 4.21 µm or less for central corneal thickness (CCT), 0.03 mm or less for anterior chamber depth (ACD) and aqueous depth (AQD), and 0.25 diopters (D) or less for mean keratometry (Km), J0, and J45. The inter-device differences in Km, J0, and J45 were statistically insignificant, whereas the remaining were statistically but not clinically significant. The 95% limits of agreement of CCT, ACD, AQD, Km, J0, and J45 were -3.70 to 15.25 µm, -0.06 to 0.04 mm, -0.06 to 0.04 mm, -0.28 to 0.35 D, -0.27 to 0.26 D, and -0.27 to 0.21 D, respectively. The double-angle plot confirmed the high agreement in astigmatism.

CONCLUSIONS

For CCT, ACD, AQD, Km, and astigmatism measurements in patients with cataract, the new SD-OCT/Placido tomographer has excellent precision and high agreement with the Argos SS-OCT biometer, and can be used interchangeably. [J Refract Surg. 2021;37(9):616-622.].

Для цитирования: Wang Qinmei, Chen Min, Ning Rui, Savini Giacomo, Wang Yiran, Zhang Tianjiao, Lin Xuanqiao, Chen Xin, Huang Jinhai. The Precision of a New Anterior Segment Optical Coherence Tomographer and Its Comparison With a Swept-Source OCT-Based Optical Biometer in Patients With Cataract.. Journal of refractive surgery (Thorofare, N.J. : 1995). 2021. DOI: 10.3928/1081597x-20210610-02.

Предметные рубрики MeSH

Подготовка и перевод публикации
Обработка статьи · этапы, ошибки и управление
Загрузка состояния…

История · было → стало

Состояние обновляется каждые 5 секунд, пока блок открыт. Процент показывается только там, где известен объём работы. Пауза останавливает продолжение этапа; текущий запрос к источнику может завершиться.

Тип публикации указан по данным источника. Уровень доказательности автоматически не присваивается.