MedElevator Войти в исследования
← Поиск публикаций

DOI 10.1177/11206721211035617

Медицинская библиотека · European journal of ophthalmology · 2021

Eight-year follow-up of Artiflex and Artiflex Toric phakic intraocular lens.

Royo Mariano, Jiménez Ángel, Martínez-Alberquilla Irene, Alfonso José F

СтатьяПеревод пока не подготовлен

Аннотация

PURPOSE: To analyse long-term efficacy, safety, visual and refractive stability and physiological changes of Artiflex Myopia and Toric phakic intraocular lenses (pIOL) separately throughout an 8-year follow-up.

DESIGN

Retrospective cohort study.

METHODS

A total of 67 eyes of 37 patients underwent Artiflex Myopia (47 eyes) or Artiflex Toric (20 eyes) implantation for correcting myopia and/or astigmatism. Follow-up evaluations were performed 1, 3, 5 and 8 years after surgery. Preoperative and postoperative data included corrected (CDVA) and uncorrected distance visual acuity (UDVA), manifest refraction, endothelial cell density (ECD) and intraocular pressure (IOP) assessments. Efficacy and safety indexes were analysed. The vectorial analysis was performed using the Thibos method.

RESULTS

Mean CDVA and UDVA of both pIOLs significantly improved from preoperative to 1 year after implantation, and then it remained stable over the 8-year follow-up. The efficacy and safety indexes after 8 years were 0.94 ± 0.16 and 1.07 ± 0.18 for Artiflex Myopia and 1.00 ± 0.11 and 1.10 ± 0.15 for Artiflex Toric, respectively. The spherical equivalent (SE) significantly improved after surgery. J0, J45 and SE refractive components showed no changes between postoperative visits. A total ECD loss of 4.8% (p < 0.001) and 10.4% (p = 0.005) was found after 8 years for Artiflex Myopia and Toric, respectively.

CONCLUSIONS

Artiflex Myopia and Toric pIOLs are a safe, efficient and predictable option for the correction of myopia and/or astigmatism. The vectorial analysis showed excellent rotation stability for the toric version.

Для цитирования: Royo Mariano, Jiménez Ángel, Martínez-Alberquilla Irene, Alfonso José F. Eight-year follow-up of Artiflex and Artiflex Toric phakic intraocular lens.. European journal of ophthalmology. 2021. DOI: 10.1177/11206721211035617.

Предметные рубрики MeSH

Подготовка и перевод публикации
Обработка статьи · этапы, ошибки и управление
Загрузка состояния…

История · было → стало

Состояние обновляется каждые 5 секунд, пока блок открыт. Процент показывается только там, где известен объём работы. Пауза останавливает продолжение этапа; текущий запрос к источнику может завершиться.

Тип публикации указан по данным источника. Уровень доказательности автоматически не присваивается.