MedElevator Войти в исследования
← Поиск публикаций

DOI 10.1177/11297298211002572

Медицинская библиотека · The journal of vascular access · 2021

ECG-guided PICC insertion using a new silicon catheter with a conductive tip: A retrospective clinical study.

Yin Yu-Xia, Gao Wei, Feng Sheng-Yu, Wang Deng-Xu, Wan Min, Liu Cheng-Hu, Wang Lu-Ning, Cao Ming-Kun, Zhang Hai-Jun

СтатьяПеревод пока не подготовлен

Аннотация

OBJECTIVE

Safety and efficacy of ECG-guided PICC insertion using a new silicon catheter with a conductive tip was evaluated in daily practice.

METHODS

A retrospective study was conducted on 1659 patients who accepted successful tip-conductive PICC placement and clinically followed-up until the catheter removal between January 2018 and April 2019. Baseline of patient characteristics, catheter placement characteristics, date of dressing changes as well as records of catheter-related complications were extracted from a special designed mobile APP.

RESULTS

The first-attempt success (success of placing catheter tip to the ideal position by primary indwelling operation) rate of PICC placement was 99.3%. The average duration of PICC placement was 128.7 ± 39.5 days and 1535 patients (92.5%) reached the therapy end-point without any complications and removed the catheter normally. The cumulative rates of total complications were 7.5%, including exit site infection (2.5%), phlebitis (0.9%), DVT (1.0%), catheter malposition (1.1%), catheter breakage (0.1%), and liquid extravasation (1.8%). In multivariable logistic regression analyses, hyperlipidemia, diabetes mellitus, lung cancer, stomach cancer, and lymphoma were significantly associated with increased risk of complications, as the independent risk factors.

CONCLUSIONS

This retrospective clinical study demonstrates that ECG-guided insertion of a new tip-conductive PICC is associated with a high rate of first-attempt success and low rate of catheter related complications.

Для цитирования: Yin Yu-Xia, Gao Wei, Feng Sheng-Yu, Wang Deng-Xu, Wan Min, Liu Cheng-Hu, Wang Lu-Ning, Cao Ming-Kun, Zhang Hai-Jun. ECG-guided PICC insertion using a new silicon catheter with a conductive tip: A retrospective clinical study.. The journal of vascular access. 2021. DOI: 10.1177/11297298211002572.

Предметные рубрики MeSH

Подготовка и перевод публикации
Обработка статьи · этапы, ошибки и управление
Загрузка состояния…

История · было → стало

Состояние обновляется каждые 5 секунд, пока блок открыт. Процент показывается только там, где известен объём работы. Пауза останавливает продолжение этапа; текущий запрос к источнику может завершиться.

Тип публикации указан по данным источника. Уровень доказательности автоматически не присваивается.