MedElevator Войти в исследования
← Поиск публикаций

DOI 10.1007/s10396-020-01013-x

Медицинская библиотека · Journal of medical ultrasonics (2001) · 2020

Endoscopic ultrasound-guided radiofrequency ablation of porcine liver.

Maeshima Shuya, Ida Yoshiyuki, Shimizu Ryo, Kawaji Yuki, Tamura Takashi, Nuta Junya, Hatamaru Keiichi, Itonaga Masahiro, Kudo Masatoshi, Kitano Masayuki

СтатьяПеревод пока не подготовлен

Аннотация

PURPOSE: Animal studies of endoscopic ultrasound-guided radiofrequency ablation (EUS-RFA) of the liver have rarely been reported. We assessed the effectiveness and safety of EUS-RFA in pigs.

METHODS

We conducted four experiments using newly designed RFA electrodes. In the first experiment, we ablated excised liver using 19 G electrodes with active electrode tips with lengths of 1, 1.5, and 2 cm. The second experiment was performed with the same electrodes as those used in the first experiment, but with the electrodes inserted into the livers of live pigs under EUS. In the third experiment, we tested the electrodes for water permeability. In the fourth experiment, we performed EUS-RFA on live pigs, using 19 G electrodes in 7/12 pigs and 18 G electrodes in 5/12 pigs. Complications were evaluated after 7 days of survival.

RESULTS

The newly designed RFA electrodes achieved ablation of the liver. In the first experiment, the maximal sizes of the ablation areas were 27, 26, 24, and 25 mm at 10, 20, 30, and 40 W, respectively, with the 2-cm electrode. In the second experiment, the maximal vertical sizes were 22, 23, 22, and 23 mm at 10, 20, 30, and 40 W, respectively, with the 2-cm electrode. In the third experiment, the 18 G electrode had better water permeability than the 19 G electrode. In the fourth experiment, all pigs survived. Complications occurred in 1/5 (18 G electrode) and 4/7 (19 G electrode) pigs.

CONCLUSION

We performed EUS-RFA in pigs and concluded that it may be feasible to perform RFA of lesions near the stomach.

Для цитирования: Maeshima Shuya, Ida Yoshiyuki, Shimizu Ryo, Kawaji Yuki, Tamura Takashi, Nuta Junya, Hatamaru Keiichi, Itonaga Masahiro, Kudo Masatoshi, Kitano Masayuki. Endoscopic ultrasound-guided radiofrequency ablation of porcine liver.. Journal of medical ultrasonics (2001). 2020. DOI: 10.1007/s10396-020-01013-x.

Предметные рубрики MeSH

Подготовка и перевод публикации
Обработка статьи · этапы, ошибки и управление
Загрузка состояния…

История · было → стало

Состояние обновляется каждые 5 секунд, пока блок открыт. Процент показывается только там, где известен объём работы. Пауза останавливает продолжение этапа; текущий запрос к источнику может завершиться.

Тип публикации указан по данным источника. Уровень доказательности автоматически не присваивается.