Low specificity of Aspergillus spp. ELITe MGB assay, and potential risks in management of invasive aspergillosis.

Bothy Violaine, Argudín Maria A, Olive Nathalie, Barbée Cindy, Kabamba-Mukadi Benoît, Rodriguez-Villalobos Hector

СтатьяПеревод пока не подготовлен

Аннотация

OBJECTIVES

In recent years, commercial molecular tools for diagnosis of invasive aspergillosis have emerged, requiring evaluation to ensure quality. Here we assessed the specificity of Aspergillus spp.-ELITe MGB Assay a commercial assay tergeting 18S gene of Aspergillus spp.

DESIGN AND

METHODS

As part of a method validation, we evaluate the specificity of the Aspergillus spp.-ELITe MGB Assay by testing fourteen culture based samples of sequenced non-Aspergillus fungal species. The benefits of a pre-lysis treatment was evaluated in parallel on serial dilutions of an Aspergillus fumigatus strain.

RESULTS

Our findings revealed cross-reactivity in five strains using the 50 copies/mL cut-off recommended by the manufacturer, suggesting potential diagnostic errors and inappropriate management of patients. Pre-lysis treatment does not affect the limit of detection at serial dilution.

CONCLUSIONS

In conclusion, the Aspergillus spp. ELITe MGB Assay exhibits limited specificity in culture-based samples, underscoring the importance of careful utilization in laboratories. Further studies are warranted to better comprehend of the impact of this cross-reactivity on clinical samples.

Для цитирования: Bothy Violaine, Argudín Maria A, Olive Nathalie, Barbée Cindy, Kabamba-Mukadi Benoît, Rodriguez-Villalobos Hector. Low specificity of Aspergillus spp. ELITe MGB assay, and potential risks in management of invasive aspergillosis.. Practical laboratory medicine. 2024. DOI: 10.1016/j.plabm.2024.e00378.
Подготовка и перевод публикации
Обработка статьи · этапы, ошибки и управление
Загрузка состояния…

История · было → стало

Состояние обновляется каждые 5 секунд, пока блок открыт. Процент показывается только там, где известен объём работы. Пауза останавливает продолжение этапа; текущий запрос к источнику может завершиться.

Тип публикации указан по данным источника. Уровень доказательности автоматически не присваивается.