MedElevator Войти в исследования
← Поиск публикаций

DOI 10.1097/ju.0000000000003449

Медицинская библиотека · The Journal of urology · 2023

A Prospective Multi-institutional Study Using a Novel Safety Valve for the Prevention of Catheter Balloon Inflation Injury of the Urethra.

O'Connor Eabhann M, Croghan Stefanie M, Baird Olivia, Fallon John, Loughman Peter, Esoof Jibraan, Keenan Robert A, Ryan James, Manecksha Rustom, D'Arcy Frank, Purtill Helen, Varma Raghav, Thursby Helen, Matkowski Amber, El Hadi Sami, Mahmalji Wasim, Giri Subhasis K

Многоцентровое исследованиеСтатьяПоддержка исследования: вне правительства СШАПеревод пока не подготовлен

Аннотация

PURPOSE: We prospectively assessed the ability of a novel transurethral catheterization safety valve to prevent urethral catheter balloon injury in a multi-institutional clinical setting.

MATERIALS AND

METHODS

A prospective, multi-institution study was conducted. The safety valve was introduced for urinary catheterization in 6 hospital groups (4 in Ireland; 2 in the UK). The safety valve allows fluid in the catheter system to vent through a pressure relief valve if attempted intraurethral inflation of the catheter's anchoring balloon occurs. Device usage was studied over a 12-month period, with data recorded using a 7-item data sticker containing a scannable QR code. "Venting" through the safety valve during catheterization was indicative of prevention of a urethral injury. An embedded 3-month study was conducted in 3 centers, with any catheter balloon injuries occurring during catheterization without safety valve use referred to the on-call urology team recorded. Health economic analyses were also performed.

RESULTS

During the overall 12-month device study phase, 994 urethral catheterizations were performed across study sites. Twenty-two (2.2%) episodes of safety valve venting were recorded. No urethral injuries occurred in these patients. In the embedded 3-month study, 18 catheter balloon injuries were recorded in association with catheterizations performed without the safety valve. Based on confirmed and device-prevented urethral injuries, the injury rate for urethral catheterization without safety valve use was calculated to be 5.5/1,000 catheterizations.

CONCLUSIONS

The safety valve has the potential to eliminate catheter balloon injury if widely adopted. It represents a simple, effective, and innovative solution to this recurring problem applicable to all patient cohorts.

Для цитирования: O'Connor Eabhann M, Croghan Stefanie M, Baird Olivia, Fallon John, Loughman Peter, Esoof Jibraan, Keenan Robert A, Ryan James, Manecksha Rustom, D'Arcy Frank, Purtill Helen, Varma Raghav, Thursby Helen, Matkowski Amber, El Hadi Sami, Mahmalji Wasim, Giri Subhasis K. A Prospective Multi-institutional Study Using a Novel Safety Valve for the Prevention of Catheter Balloon Inflation Injury of the Urethra.. The Journal of urology. 2023. DOI: 10.1097/ju.0000000000003449.

Предметные рубрики MeSH

Подготовка и перевод публикации
Обработка статьи · этапы, ошибки и управление
Загрузка состояния…

История · было → стало

Состояние обновляется каждые 5 секунд, пока блок открыт. Процент показывается только там, где известен объём работы. Пауза останавливает продолжение этапа; текущий запрос к источнику может завершиться.

Тип публикации указан по данным источника. Уровень доказательности автоматически не присваивается.