A new urinary catheter design reduces in-vitro biofilm formation by influencing hydrodynamics.
СтатьяНазвание и аннотация: Перевод пока не подготовлен
Обработка статьи · этапы, ошибки и управление
Загрузка состояния…
Ошибки и ограничения
История · было → стало
Состояние обновляется каждые 5 секунд, пока блок открыт. Процент показывается только там, где известен объём работы. Пауза останавливает продолжение этапа; текущий запрос к источнику может завершиться.
Ionescu A C, Brambilla E, Sighinolfi M C, Mattina R
Аннотация
AIM: To evaluate the performance of a new catheter design based on different hydrodynamics aiming to reduce the development of biofilm, and compare it with a conventional Foley catheter (FC).
METHODS: The new proposed design (NPD) catheter is a modification of the FC, based on asymmetric positioning of the balloon and additional drainage holes allowing continuous urine drainage and complete voiding of the bladder. A first experiment was undertaken to assess drainage capability, and a second experiment was performed using a bioreactor with a set-up simulating the bladder and using the test catheter as a flow-through system. The biofilm formation of five bacterial species associated with catheter-associated urinary tract infection (CAUTI) was determined after 24 h of incubation using an MTT assay. Morphological evaluation was performed using scanning electron microscopy. In-vitro determination of residual fluid, and quantitative and morphological data on biofilm formation on the intravesical and intraluminal parts of the tested catheters were assessed.
RESULTS: Residual fluid was significantly higher in the FC (5.60 ± 0.43 mL) compared with the NPD catheter (0.2 ± 0.03 mL). The NPD catheter showed significantly less biofilm formation (P<0.0001) than the FC. Catheter design had a variable effect on biofilm formation depending on the bacterial strain tested. There was significantly less intraluminal biomass compared with intravesical biomass in both catheters (P<0.0001). Multi-layered biofilms that covered the FC surfaces completely were seen for all tested strains, while the NPD catheter surfaces showed reduced biofilm formation.
CONCLUSIONS: Modifications of the hydrodynamic characteristics of a catheter can significantly reduce bacterial colonization. Integrated design approaches combining chemical, mechanical and topographical elements can help to reduce the occurrence of CAUTI.
METHODS: The new proposed design (NPD) catheter is a modification of the FC, based on asymmetric positioning of the balloon and additional drainage holes allowing continuous urine drainage and complete voiding of the bladder. A first experiment was undertaken to assess drainage capability, and a second experiment was performed using a bioreactor with a set-up simulating the bladder and using the test catheter as a flow-through system. The biofilm formation of five bacterial species associated with catheter-associated urinary tract infection (CAUTI) was determined after 24 h of incubation using an MTT assay. Morphological evaluation was performed using scanning electron microscopy. In-vitro determination of residual fluid, and quantitative and morphological data on biofilm formation on the intravesical and intraluminal parts of the tested catheters were assessed.
RESULTS: Residual fluid was significantly higher in the FC (5.60 ± 0.43 mL) compared with the NPD catheter (0.2 ± 0.03 mL). The NPD catheter showed significantly less biofilm formation (P<0.0001) than the FC. Catheter design had a variable effect on biofilm formation depending on the bacterial strain tested. There was significantly less intraluminal biomass compared with intravesical biomass in both catheters (P<0.0001). Multi-layered biofilms that covered the FC surfaces completely were seen for all tested strains, while the NPD catheter surfaces showed reduced biofilm formation.
CONCLUSIONS: Modifications of the hydrodynamic characteristics of a catheter can significantly reduce bacterial colonization. Integrated design approaches combining chemical, mechanical and topographical elements can help to reduce the occurrence of CAUTI.