MedElevator Войти в исследования
← Поиск публикаций

DOI 10.1109/tbme.2021.3066995

Медицинская библиотека · IEEE transactions on bio-medical engineering · 2021

Integrated System for Bacterial Detection and Biofilm Treatment on Indwelling Urinary Catheters.

Huiszoon Ryan C, Han Jinjing, Chu Sangwook, Stine Justin M, Beardslee Luke A, Ghodssi Reza

СтатьяПоддержка исследования: вне правительства СШАПоддержка правительства СШАПеревод пока не подготовлен

Аннотация

GOAL: This work introduces an integrated system incorporated seamlessly with a commercial Foley urinary catheter for bacterial growth sensing and biofilm treatment.

METHODS

The system is comprised of flexible, interdigitated electrodes incorporated with a urinary catheter via a 3D-printed insert for impedance sensing and bioelectric effect-based treatment. Each of the functions were wirelessly controlled using a custom application that provides a user-friendly interface for communicating with a custom PCB via Bluetooth to facilitate implementation in practice.

RESULTS

The integrated catheter system maintains the primary functions of indwelling catheters - urine drainage, balloon inflation - while being capable of detecting the growth of Escherichia coli, with an average decrease in impedance of 13.0% after 24 hours, tested in a newly-developed simulated bladder environment. Furthermore, the system enables bioelectric effect-based biofilm reduction, which is performed by applying a low-intensity electric field that increases the susceptibility of biofilm bacteria to antimicrobials, ultimately reducing the required antibiotic dosage.

CONCLUSION

Overall, this modified catheter system represents a significant step forward for catheter-associated urinary tract infection (CAUTI) management using device-based approaches, integrating flexible electrodes with an actual Foley catheter along with the control electronics and mobile application.

SIGNIFICANCE: CAUTIs, exacerbated by the emergence of antibiotic-resistant pathogens, represent a significant challenge as one of the most prevalent healthcare-acquired infections. These infections are driven by the colonization of indwelling catheters by bacterial biofilms.

Для цитирования: Huiszoon Ryan C, Han Jinjing, Chu Sangwook, Stine Justin M, Beardslee Luke A, Ghodssi Reza. Integrated System for Bacterial Detection and Biofilm Treatment on Indwelling Urinary Catheters.. IEEE transactions on bio-medical engineering. 2021. DOI: 10.1109/tbme.2021.3066995.

Предметные рубрики MeSH

Подготовка и перевод публикации
Обработка статьи · этапы, ошибки и управление
Загрузка состояния…

История · было → стало

Состояние обновляется каждые 5 секунд, пока блок открыт. Процент показывается только там, где известен объём работы. Пауза останавливает продолжение этапа; текущий запрос к источнику может завершиться.

Тип публикации указан по данным источника. Уровень доказательности автоматически не присваивается.