DOI 10.1016/j.jcrs.2012.11.019

Медицинская библиотека · Journal of cataract and refractive surgery · 2013

Anterior chamber depth, intraocular lens position, and refractive outcomes after cataract surgery.

Engren Anna-Lotta, Behndig Anders

Сравнительное исследованиеСтатьяПеревод пока не подготовлен

Аннотация

PURPOSE: To assess postoperative intraocular lens (IOL) position in relation to the cornea and iris, compare 2 devices for anterior chamber depth (ACD) measurement, and assess the impact of these factors on postoperative refraction in phacoemulsification.

SETTING

Department of Clinical Sciences/Ophthalmology, Umeå University Hospital, Umeå, Sweden.

DESIGN

Evaluation of diagnostic test or technology.

METHODS

Biometry and measurement of the anterior segment were performed preoperatively and postoperatively with the partial coherence interferometry (PCI)-based IOLMaster and the Pentacam HR Scheimpflug device. Predicted refraction was calculated with ACD estimations according to the Haigis formula and was compared with the actual postoperative refraction. Changes in ACD and the distance between the cornea-iris and the iris-lens/IOL were assessed from Scheimpflug images. The theoretical refractive impact of differences in ACD was calculated.

RESULTS

The mean preoperative ACD and refractive prediction error, respectively, was 3.22 mm ± 0.37 (SD) and 0.41 ± 0.43 diopters (D) for the PCI device and 3.21 ± 0.35 mm and 0.41 ± 0.41 D for the Scheimpflug device; the difference was not statistically significant. The mean ACD, cornea-iris distance, and iris-lens/IOL distance changes were 1.62 ± 0.38 mm, 0.78 ± 0.29 mm, and 0.85 ± 0.27 mm, respectively. The mean predicted refractive outcome from a 1.0 mm difference in ACD was 0.32 D.

CONCLUSIONS

No difference was found between the measurement devices. Although the exact postoperative IOL position is difficult to predict, its impact on postoperative refraction was comparatively small compared with the impact of minor corneal curvature or axial length measurement errors.

Для цитирования: Engren Anna-Lotta, Behndig Anders. Anterior chamber depth, intraocular lens position, and refractive outcomes after cataract surgery.. Journal of cataract and refractive surgery. 2013. DOI: 10.1016/j.jcrs.2012.11.019.

Предметные рубрики MeSH

Подготовка и перевод публикации
Обработка статьи · этапы, ошибки и управление
Загрузка состояния…

История · было → стало

Состояние обновляется каждые 5 секунд, пока блок открыт. Процент показывается только там, где известен объём работы. Пауза останавливает продолжение этапа; текущий запрос к источнику может завершиться.

Тип публикации указан по данным источника. Уровень доказательности автоматически не присваивается.