Correlation between ORange (Gen 1, pseudophakic) intraoperative refraction and 1-week postcataract surgery autorefraction.

Chen Ming

СтатьяПеревод пока не подготовлен

Аннотация

BACKGROUND

This study aimed to find the strength of the correlation between ORange(®) (intraoperative wavefront aberrometer) objective refraction during pseudophakic cataract surgery (Gen 1) and 1-week objective autorefraction postcataract surgery in order to establish the clinical degree of confidence in the ORange. Gen 1 was used because Gen 2 was not available at the time of the study. However, Gen 1 can still be a good reference.

METHODS

Thirty-two consecutive eyes were included in this prospective nonrandomized study. ORange was used after phacoemulsification and intraocular lens implantation. There were no complications and no withdrawals from the study. The refraction data from the ORange were recorded in spherical equivalents and put into Excel. One week postoperatively, autorefraction was done and the data were again recorded in spherical equivalents and put into Excel. Both sets of data were then transferred to SPSS for a correlation study. Because ORange is an objective refractor, it is necessary to use another objective refractor, such as the Zeiss Autorefractor, for comparison. Measurement at 1 week postoperatively is for cornea clearance for autorefraction and is the closest time to intraoperative assessment for comparison.

RESULTS

The Pearson correlation was r = +0.56, P < 0.01, which qualifies for a moderate or good correlation (0.5-0.75).

CONCLUSION

Intraoperative refraction using the ORange Gen 1 after cataract surgery and intraocular lens implantation is moderate to good when correlated with objective autorefraction 1 week after cataract surgery.

Для цитирования: Chen Ming. Correlation between ORange (Gen 1, pseudophakic) intraoperative refraction and 1-week postcataract surgery autorefraction.. Clinical ophthalmology (Auckland, N.Z.). 2011. DOI: 10.2147/opth.s17489.
Подготовка и перевод публикации
Обработка статьи · этапы, ошибки и управление
Загрузка состояния…

История · было → стало

Состояние обновляется каждые 5 секунд, пока блок открыт. Процент показывается только там, где известен объём работы. Пауза останавливает продолжение этапа; текущий запрос к источнику может завершиться.

Тип публикации указан по данным источника. Уровень доказательности автоматически не присваивается.