Journal of vascular and interventional radiology : JVIR · 2011
Retrievability and device-related complications of the G2 filter: a retrospective study of 139 filter retrievals.
СтатьяНазвание и аннотация: Перевод пока не подготовлен
Обработка статьи · этапы, ошибки и управление
Загрузка состояния…
Ошибки и ограничения
История · было → стало
Состояние обновляется каждые 5 секунд, пока блок открыт. Процент показывается только там, где известен объём работы. Пауза останавливает продолжение этапа; текущий запрос к источнику может завершиться.
Zhu Xiaoli, Tam Matthew D B S, Bartholomew John, Newman James S, Sands Mark J, Wang Weiping
Аннотация
PURPOSE: To evaluate the retrievability and safety of the G2 filter.
MATERIALS AND METHODS: A retrospective study of all G2 filter retrievals at a single institution was conducted. Hospital records and imaging studies were reviewed for complications, and factors affecting retrieval were analyzed.
RESULTS: From 2005 to 2009, a total of 139 patients presented for retrieval of their G2 filter, and 131 pairs of pre- and post-placement cavagrams and 39 computed tomography scans were available for analysis. The following findings were recorded: limb penetration (n = 33), tilt greater than 15° (n = 22), local migration greater than 2 cm (n = 17), retained thrombus within the filter (n = 16), deformity (n = 10), inferior vena cava (IVC) occlusion (n = 3), fracture (n = 2), and pulmonary embolism breakthrough (n = 2). A total of 118 filters were removed, with a mean indwelling time of 131.8 days (range, 3-602 d). Indwell time (< 90, 90-180, or > 180 d) did not affect retrieval (P = .4). There were 21 filters (15.1%) left in situ as a result of severe tilt (n = 9), significant thrombus in the filter (n = 5), IVC occlusion (n = 3), filter incorporation into the caval wall (n = 3), or lack of central venous access (n = 1). There was a strong relationship between penetration and caudal migration (P < .0001). Severe tilt was associated with prolonged fluoroscopic times for retrieval (P = .003).
CONCLUSIONS: The majority of G2 filters can be removed without difficulty. The most common factor affecting retrieval was severe tilting. The indwelling time had no impact on retrieval. G2 filter-related complications were frequent but most, including fractures, were clinically insignificant.
MATERIALS AND METHODS: A retrospective study of all G2 filter retrievals at a single institution was conducted. Hospital records and imaging studies were reviewed for complications, and factors affecting retrieval were analyzed.
RESULTS: From 2005 to 2009, a total of 139 patients presented for retrieval of their G2 filter, and 131 pairs of pre- and post-placement cavagrams and 39 computed tomography scans were available for analysis. The following findings were recorded: limb penetration (n = 33), tilt greater than 15° (n = 22), local migration greater than 2 cm (n = 17), retained thrombus within the filter (n = 16), deformity (n = 10), inferior vena cava (IVC) occlusion (n = 3), fracture (n = 2), and pulmonary embolism breakthrough (n = 2). A total of 118 filters were removed, with a mean indwelling time of 131.8 days (range, 3-602 d). Indwell time (< 90, 90-180, or > 180 d) did not affect retrieval (P = .4). There were 21 filters (15.1%) left in situ as a result of severe tilt (n = 9), significant thrombus in the filter (n = 5), IVC occlusion (n = 3), filter incorporation into the caval wall (n = 3), or lack of central venous access (n = 1). There was a strong relationship between penetration and caudal migration (P < .0001). Severe tilt was associated with prolonged fluoroscopic times for retrieval (P = .003).
CONCLUSIONS: The majority of G2 filters can be removed without difficulty. The most common factor affecting retrieval was severe tilting. The indwelling time had no impact on retrieval. G2 filter-related complications were frequent but most, including fractures, were clinically insignificant.
Предметные рубрики MeSH
AdultAgedAged, 80 and overChi-Square DistributionChinaDevice RemovalFemaleForeign-Body MigrationHumansMaleMiddle AgedPhlebographyProsthesis DesignProsthesis FailurePulmonary EmbolismRetrospective StudiesRisk AssessmentRisk FactorsTime FactorsTomography, X-Ray ComputedTreatment OutcomeVena Cava FiltersVena Cava, InferiorVenous ThrombosisYoung Adult