DOI 10.1016/j.jvir.2009.03.046
G2 inferior vena cava filter: retrievability and safety.
Аннотация
PURPOSE: To assess the retrievability of the G2 inferior vena cava (IVC) filter and factors influencing the safety and technical success of retrieval.
MATERIALS AND
METHODS
From October 2006 through June 2008, G2 IVC filters were placed in 140 consecutive patients who needed prophylaxis against pulmonary embolism (PE). General indications for filter placement included history of thromboembolic disease (n = 98) and high risk for PE (n = 42); specific indications included contraindication to anticoagulation (n = 120), prophylaxis in addition to anticoagulation (n = 16), and failure of anticoagulation (n = 4). Filter dwell time, technical success of filter retrieval, and complications related to placement or retrieval were retrospectively evaluated in patients who underwent filter removal.
RESULTS
Twenty-seven attempts at G2 filter removal were made in 26 patients (12 men; age range, 24-88 years; mean age, 55.4 y) after a mean period of 122 days (range, 11-260 d). Data were collected retrospectively with institutional review board approval. Filter removal was successful in all 27 attempts (100%). Tilting of the filter (> or =15 degrees ) occurred in five cases (18.5%), with probable filter incorporation into the right lateral wall of the IVC in one. Other complications of retrieval such as filter thrombosis, significant filter migration, filter fracture, and caval occlusion were not observed.
CONCLUSIONS
G2 IVC filter retrieval has a high technical success rate and a low complication rate. Technical success appears to be unaffected by the dwell time within the reported range.
Предметные рубрики MeSH
Подготовка и перевод публикации
Обработка статьи · этапы, ошибки и управление
Ошибки и ограничения
История · было → стало
Состояние обновляется каждые 5 секунд, пока блок открыт. Процент показывается только там, где известен объём работы. Пауза останавливает продолжение этапа; текущий запрос к источнику может завершиться.
Тип публикации указан по данным источника. Уровень доказательности автоматически не присваивается.