DOI 10.1097/j.jcrs.0000000000001735

Медицинская библиотека · Journal of cataract and refractive surgery · 2025

Variations in morphology and patency of disposable hydrodissection cannulas used in cataract surgery.

Mudalegundi Shwetha, Yao Melissa, Chew Jesse, Kelkar Neil, Eid Kevin, Perez Jesus Paulo, Driffill Ethan, Mamalis Nick, Werner Liliana

СтатьяПеревод пока не подготовлен

Аннотация

PURPOSE: To perform a detailed laboratory assessment of the variability in morphology (shape of opening port and surface finishing) and patency of Chang-style hydrodissection cannulas used in cataract surgery.

SETTING

John A. Moran Eye Center, University of Utah, Salt Lake City, Utah.

DESIGN

Experimental study.

METHODS

44 disposable cannulas from 3 different manufacturers were evaluated in this study. 17 cannulas were primed with solution and then underwent analyses under gross and light microscopy. 27 other cannulas were directly removed from the package and underwent the same abovementioned analyses, before priming. Selected cannulas from the 3 manufacturers were also evaluated under scanning electron microscopy and surface analyses for elemental composition.

RESULTS

There was variability in all parameters evaluated in this study, including opening ports that were irregular in shape, partially or totally blocked, and a cannula with a second, lateral opening port. The surface finishing of the tips of the cannulas was also variable, with cannulas exhibiting irregular and/or sharp edges, as well as what seemed to be debris/residues. Fluid streams obtained with the cannulas also varied, and although some cannulas produced a normal stream, others produced a weak stream, no stream, or double stream. All cannulas were composed of stainless steel, and debris/residues seemed to correspond to organic residues.

CONCLUSIONS

The findings in this study raise concerns about potential damage to the capsule and toxic anterior segment syndrome. Priming before use in surgery is advisable to confirm appropriate fluid flow.

Для цитирования: Mudalegundi Shwetha, Yao Melissa, Chew Jesse, Kelkar Neil, Eid Kevin, Perez Jesus Paulo, Driffill Ethan, Mamalis Nick, Werner Liliana. Variations in morphology and patency of disposable hydrodissection cannulas used in cataract surgery.. Journal of cataract and refractive surgery. 2025. DOI: 10.1097/j.jcrs.0000000000001735.

Предметные рубрики MeSH

Подготовка и перевод публикации
Обработка статьи · этапы, ошибки и управление
Загрузка состояния…

История · было → стало

Состояние обновляется каждые 5 секунд, пока блок открыт. Процент показывается только там, где известен объём работы. Пауза останавливает продолжение этапа; текущий запрос к источнику может завершиться.

Тип публикации указан по данным источника. Уровень доказательности автоматически не присваивается.