Performance and usability evaluation of three LDH-based malaria rapid diagnostic tests in Kédougou, Senegal.

Sambe Babacar Souleymane, Zobrist Stephanie, Sheahan William, Soni Divya, Diagne Aissatou, Sarr Ibrahima, Diatta Arona Sabene, Mbacke Diaw Serigne Ousmane, Golden Allison, Slater Hannah, Jang Ihn Kyung, Roa Nerie, Pal Sampa, Sarr Fatoumata Diene, Faye Joseph, Vigan-Womas Inès, Dieye Yakou, Cisse Moustapha, Domingo Gonzalo J, Niang Makhtar

СтатьяПрепринтПеревод пока не подготовлен

Аннотация

BACKGROUND

The emergence of pfhrp2/3-deleted parasites threatens histidine-rich protein 2 (HRP2)-based malaria rapid diagnostic test (RDT) performance. RDTs targeting Plasmodium falciparum (Pf) lactate dehydrogenase (LDH) may address current product limitations and improve case management.

OBJECTIVES

To evaluate the performance and usability of three LDH-based RDTs in febrile patients.

METHODS

A cross-sectional diagnostic accuracy study was conducted in Kédougou, Senegal. Capillary blood was tested using the SD Bioline Ag Pf (#05FK50) and three LDH-based RDTs: BIOCREDIT Malaria Ag Pf (pLDH), BIOCREDIT Pf (pLDH/HRPII), and BIOCREDIT Pf/Pv (pLDH/pLDH) (Rapigen Inc., Republic of Korea). Venous blood was collected to repeat the BIOCREDIT RDTs and conduct microscopy. Frozen venous specimens were tested using a reference PCR assay. A quantitative multiplex malaria antigen assay measured antigen concentration. RDT performance was determined and analyzed as a function of antigen concentration distribution. Usability of the Pf-only BIOCREDIT tests was evaluated using a questionnaire.

RESULTS

154/220 participants (70%) were Pf-positive by PCR. The Pf (pLDH/HRPII) test demonstrated the highest sensitivity at 78% (70.9%-84.5%); specificity was 89% (79.4%-95.6%). All RDTs performed better than microscopy (53% sensitivity). RDTs performed better when compared to antigen concentration over PCR results. Improved sensitivity of the Pf (pLDH/HRPII) test was driven by the HRP2 line. Line intensity correlated with antigen concentration. Predicted sensitivity using the analytical limit of detection (LOD) was comparable to observed sensitivity. RDTs demonstrated acceptable usability.

CONCLUSIONS

Both HRP2 and LDH contributed to the sensitivity of the best-performing Pf-RDT. RDT analytical LODs can be used to predict performance in populations with known antigen concentrations.

Для цитирования: Sambe Babacar Souleymane, Zobrist Stephanie, Sheahan William, Soni Divya, Diagne Aissatou, Sarr Ibrahima, Diatta Arona Sabene, Mbacke Diaw Serigne Ousmane, Golden Allison, Slater Hannah, Jang Ihn Kyung, Roa Nerie, Pal Sampa, Sarr Fatoumata Diene, Faye Joseph, Vigan-Womas Inès, Dieye Yakou, Cisse Moustapha, Domingo Gonzalo J, Niang Makhtar. Performance and usability evaluation of three LDH-based malaria rapid diagnostic tests in Kédougou, Senegal.. medRxiv : the preprint server for health sciences. 2024. DOI: 10.1101/2024.12.12.24318945.
Подготовка и перевод публикации
Обработка статьи · этапы, ошибки и управление
Загрузка состояния…

История · было → стало

Состояние обновляется каждые 5 секунд, пока блок открыт. Процент показывается только там, где известен объём работы. Пауза останавливает продолжение этапа; текущий запрос к источнику может завершиться.

Тип публикации указан по данным источника. Уровень доказательности автоматически не присваивается.